ISO / ISO 10993-12:2021/Amd 1:2025 Biological evaluation of medical devices — Part 12: Sample preparation and reference materials — AMENDMENT 1

ISO 10993-12:2021/Amd 1:2025 Biological evaluation of medical devices — Part 12: Sample preparation and reference materials — AMENDMENT 1

Foreword

ISO (the International Organization for Standardization) is a worldwide federation of national standards bodies (ISO member bodies). The work of preparing International Standards is normally carried out through ISO technical committees. Each member body interested in a subject for which a technical committee has been established has the right to be represented on that committee. International organizations, governmental and non-governmental, in liaison with ISO, also take part in the work. ISO collaborates closely with the International Electrotechnical Commission (IEC) on all matters of electrotechnical standardization.

The procedures used to develop this document and those intended for its further maintenance are described in the ISO/IEC Directives, Part 1. In particular, the different approval criteria needed for the different types of ISO documents should be noted. This document was drafted in accordance with the editorial rules of the ISO/IEC Directives, Part 2 (see www.iso.org/directives).

ISO draws attention to the possibility that the implementation of this document may involve the use of (a) patent(s). ISO takes no position concerning the evidence, validity or applicability of any claimed patent rights in respect thereof. As of the date of publication of this document, ISO had not received notice of (a) patent(s) which may be required to implement this document. However, implementers are cautioned that this may not represent the latest information, which may be obtained from the patent database available at www.iso.org/patents. ISO shall not be held responsible for identifying any or all such patent rights.

Any trade name used in this document is information given for the convenience of users and does not constitute an endorsement.

For an explanation of the voluntary nature of standards, the meaning of ISO specific terms and expressions related to conformity assessment, as well as information about ISO's adherence to the World Trade Organization (WTO) principles in the Technical Barriers to Trade (TBT), see www.iso.org/iso/foreword.html.

This document was prepared by Technical Committee ISO/TC 194, Biological and clinical evaluation of medical devices, in collaboration with the European Committee for Standardization (CEN) Technical Committee CEN/TC 206, Biological and clinical evaluation of medical devices, in accordance with the Agreement on technical cooperation between ISO and CEN (Vienna Agreement).

A list of all parts in the ISO 10993 series can be found on the ISO website.

Any feedback or questions on this document should be directed to the user’s national standards body. A complete listing of these bodies can be found at www.iso.org/members.html.

Biological evaluation of medical devices — Part 12: Sample preparation and reference materials — AMENDMENT 1

 

Clause 2

Replace the text of Clause 2 with:

The following documents are referred to in the text in such a way that some or all of their content constitutes requirements of this document. For dated references, only the edition cited applies. For undated references, the latest edition of the referenced document (including any amendments) applies.

ISO 10993-1, Biological evaluation of medical devices — Part 1: Evaluation and testing within a risk management process

ISO 10993-18, Biological evaluation of medical devices — Part 18: Chemical characterization of medical device materials within a risk management process

ISO/TS 10993-19, Biological evaluation of medical devices — Part 19: Physico-chemical, morphological and topographical characterization of materials

ISO 14971, Medical devices — Application of risk management to medical devices

 

Clause 4

Change the wording in the first sentence of the first paragraph from “as described in ISO 14971” to “in accordance with ISO 14971”.

 

Clause 7

Change the wording in the first sentence of the first paragraph from “(see ISO 10993-1)” to “in accordance with ISO 10993-1”.

 

10.1

Change the wording in the second sentence of the first paragraph from “(see ISO 10993-18 and ISO/TS 10993-19)” to “in accordance with ISO 10993-18 and ISO/TS 10993-19”.

 

10.3.3, first paragraph

Add the following paragraph after the first paragraph:

For medical devices containing absorbent materials, the individual absorption capacity of these materials should be considered to determine the overall extraction volume (see Table 1).

 

10.3.3, Table 1

Replace Table 1 with:

Table 1
Standard surface areas and extract liquid volumes
Thicknessa

mm

Extraction ratio
(surface area or mass, per volume)

±10 %

Examples of forms of materials
<0,5 6 cm2/ml film, sheet, tubing wall
0,5 to 1,0 3 cm2/ml tubing wall, slab, small moulded items
>1,0 3 cm2/ml larger moulded items
>1,0 1,25 cm2/ml elastomeric materials
irregularly shaped solid devices 0,2 g/ml powder, pellets, foam, non-absorbent moulded items, porous high-density materials
irregularly shaped porous devices
(low-density materials)
0,1 g/ml membranes, textiles

NOTE While there are no standardized methods available at present for testing solvent absorbing polymer materials
(e.g. absorbents and hydrocolloids), a suggested protocol is as follows:

—   first, the volume of extraction vehicle that each 0,1 g or 1,0 cm2 of material absorbs is determined;

—   then, in performing the material extraction, the additional volume is added to each 0,1 g or 1,0 cm2 in an extraction mixture.

a If the medical device includes multiple tissue contacting components with different thicknesses, the extraction ratio should be justified. One way to do this is to base the ratio on the thinnest material layer of that component.

 

10.3.4

Delete the last sentence: "For material such as elastomers and latex, it may be appropriate to use an extraction ratio of 1,25 cm2/ml, if justified".