| 5.1 Manufacture |
| 5.1.1 |
Manufacturer |
Indicates the medical device manufacturer |
This symbol shall be accompanied by the name and address of the manufacturer adjacent to the symbol. |
NOTE 1 This symbol is used to indicate information that is required in Europe and can be required by other authorities having jurisdiction. NOTE 2 For use in Europe the full definition of “manufacturer” is given in EU Regulations 2017/745[23] and 2017/746.[24] Other jurisdictions can have unique definitions. NOTE 3 The date of manufacture, as well as the name and address of the manufacturer, can be combined in one symbol. |
— |
ISO 7000-3082 2011-10-02 |
 |
| 5.1.2 |
Authorized representative in the European Community/ European Union |
Indicates the authorized representative in the European Community/ European Union |
This symbol shall be accompanied by the name and address of the authorized representative, adjacent to the symbol. |
NOTE 1 This symbol is used to indicate information that is required in the European Community/European Union. NOTE 2 Additional guidance can be found in ISO 20417[15], ISO 18113-1[10], ISO 18113-2[11], ISO 18113-3[12], ISO 18113-4[13] and ISO 18113-5[14]. NOTE 3 If multiple symbols (i.e. Authorized Representative, Importer, Distributor, Translation, or Repackaging) identify the same responsible entity, the name and address need not be duplicated, and all applicable symbols can be grouped together next to the single address. |
— |
N/A |
 |
| 5.1.3 |
Date of manufacture |
Indicates the date when the medical device was manufactured |
This symbol shall be accompanied by a date to indicate the date of manufacture. This shall be expressed in accordance with ISO 8601-1. The date shall be located adjacent to the symbol. |
— |
The use of this symbol precludes the use of symbol 5.1.11 with a date of manufacture. |
ISO 7000-2497 2004-01-15 |
 |
| 5.1.4 |
Use-by date |
Indicates the date after which the medical device is not to be used |
This symbol shall be accompanied by a date to indicate that the medical device should not be used after the end of the year, month or day shown. The date shall be expressed in accordance with ISO 8601-1. The date shall be located adjacent to the symbol. |
NOTE Synonyms for “use-by date” are “use by”, “expiry date” and “expiration date”. |
— |
ISO 7000-2607 2004-01-15 |
 |
| 5.1.5 |
Batch code |
Indicates the manufacturer’s batch code so that the batch or lot can be identified |
This symbol shall be accompanied by the manufacturer’s batch code adjacent to the symbol. |
NOTE Synonyms for “batch code” are “lot number”, “lot code” and “batch number”. |
— |
ISO 7000-2492 2004-01-15 |
 |
| 5.1.6 |
Catalogue number |
Indicates the manufacturer’s catalogue number so that the medical device can be identified |
This symbol shall be accompanied by the manufacturer’s catalogue number adjacent to the symbol. |
NOTE Synonyms for “catalogue number” are “commercial product name”, “commercial product code”, stock keeping unit, “reference number” and “reorder number”. |
— |
ISO 7000-2493 2004-01-15 |
 |
| 5.1.7 |
Serial number |
Indicates the manufacturer’s serial number so that a specific medical device can be identified |
This symbol shall be accompanied by the manufacturer’s serial number adjacent to the symbol. |
— |
— |
ISO 7000-2498 2004-01-15 |
 |
| 5.1.8 |
Importer |
Indicates the entity importing the medical device into the locale |
This symbol shall be accompanied by the name and address of the importing entity adjacent to the symbol. |
NOTE If multiple symbols (i.e. Authorized Representative, Importer, Distributor, Translation, or Repackaging) identify the same responsible entity, the name and address need not be duplicated. |
— |
ISO 7000-3725 2019-11-01 |
 |
| 5.1.9 |
Distributor |
Indicates the entity distributing the medical device into the locale |
This symbol shall be accompanied by the name and address of the distributing entity adjacent to the symbol. |
NOTE If multiple symbols (i.e. Authorized Representative, Importer, Distributor, Translation, or Repackaging) identify the same responsible entity, the name and address need not be duplicated. |
— |
ISO 7000-3724 2019-11-01 |
 |
| 5.1.10 |
Model number |
Indicates the model number or type number of a product |
This symbol shall be accompanied by the model number of the product adjacent to the symbol. |
— |
— |
IEC 60417-6050 2012-07-14 |
 |
| 5.1.11 |
Country of manufacture |
To identify the country of manufacture of products |
In the application of this symbol, the “CC” shall be replaced by either the two-letter country code or the three letter country code defined in ISO 3166-1.The date of manufacture may be added adjacent to this symbol. |
NOTE Not all authorities with jurisdiction recognize the two letter or three letter country codes found in ISO 3166-1. |
The use of this symbol with a date of manufacture precludes the use of symbol 5.1.3. |
IEC 60417-6049 2012-07-14 |
 |
| 5.2 Sterility |
| 5.2.1 |
Sterile |
Indicates a medical device that has been subjected to a sterilization process |
— |
— |
Use of this symbol precludes the use of symbols 5.2.2 to 5.2.5 or 5.2.10. |
ISO 7000-2499 2004-01-15 |
 |
| 5.2.2 |
Sterilized using aseptic processing techniques |
Indicates a medical device that has been manufactured using accepted aseptic techniques |
— |
NOTE Aseptic techniques can include filtration. |
Use of this symbol precludes the use of symbol 5.2.1. |
ISO 7000-2500 2004-01-15 |
 |
| 5.2.3 |
Sterilized using ethylene oxide |
Indicates a medical device that has been sterilized using ethylene oxide |
— |
— |
Use of this symbol precludes the use of symbol 5.2.1. |
ISO 7000-2501 2004-01-15 |
 |
| 5.2.4 |
Sterilized using irradiation |
Indicates a medical device that has been sterilized using irradiation |
— |
— |
Use of this symbol precludes the use of symbol 5.2.1. |
ISO 7000-2502 2004-01-15 |
 |
| 5.2.5 |
Sterilized using steam or dry heat |
Indicates a medical device that has been sterilized using steam or dry heat |
— |
— |
Use of this symbol precludes the use of symbol 5.2.1. |
ISO 7000-2503 2004-01-15 |
 |
| 5.2.6 |
Do not resterilize |
Indicates a medical device that is not to be resterilized |
— |
— |
This symbol is only to be used when there is an accompanying Sterile symbol (5.2.1 to 5.2.5 or 5.2.10). This symbol is not to be used on reusable medical devices that are intended to be sterilized between uses. |
ISO 7000-2608 2004-01-15 |
 |
| 5.2.7 |
Non-sterile |
Indicates a medical device that has not been subjected to a sterilization process |
— |
— |
This symbol should only be used to distinguish between identical or similar medical devices sold in both sterile and non-sterile conditions.Use of this symbol precludes the use of symbols 5.2.1 to 5.2.5 and 5.2.10. |
ISO 7000-2609 2004-01-15 |
 |
| 5.2.8 |
Do not use if package is damaged and consult instructions for use |
Indicates that a medical device that should not be used if the package has been damaged or opened and that the user should consult the instructions for use for additional information |
— |
NOTE 1 This symbol can also mean “Do not use if the product sterile barrier system or its packaging is compromised”.NOTE 2 For products that do not have instructions for use, the recommendation to consult them does not apply. |
— |
ISO 7000-2606 2004-01-15 |
 |
| 5.2.9 |
Sterile fluid path |
Indicates the presence of a sterile fluid path within the medical device in cases when other parts of the medical device, including the exterior, might not be supplied sterile |
The method of sterilization shall be indicated in the empty portion of the symbol, as appropriate. The part of the medical device that is sterile shall be identified in the information supplied by the manufacturer. |
— |
— |
ISO 7000-3084 2011-10-05 |
 |
| 5.2.10 |
Sterilized using vaporized hydrogen peroxide |
Indicates a medical device that has been sterilized using vaporized hydrogen peroxide |
— |
NOTE The use of this symbol in Europe is explained in EN 556-1:2001,[21] 4.1 and the associated note. |
Use of this symbol precludes the use of symbol 5.2.1. |
N/A |
 |
| 5.2.11 |
Single sterile barrier system |
Indicates a single sterile barrier system |
This symbol shall be placed adjacent to or in combination with symbol 5.2.1, 5.2.2, 5.2.3, 5.2.4, 5.2.5, 5.2.9 or 5.2.10. |
NOTE 1 A solid line identifies a sterile barrier system. NOTE 2 Additional information on sterile barrier systems can be found in ISO 11607-1[5] and ISO 11607-2[6]. |
— |
ISO 7000-3707 2019-10-18 |
 |
| 5.2.12 |
Double sterile barrier system |
Indicates two sterile barrier systems |
This symbol shall be placed adjacent to or in combination with symbol 5.2.1, 5.2.2, 5.2.3, 5.2.4, 5.2.5, 5.2.9 or 5.2.10. |
NOTE 1 A double solid line indicates a double sterile barrier system. NOTE 2 Additional information on sterile barrier systems can be found in ISO 11607-1[5] and ISO 11607-2[6]. |
— |
ISO 7000-3704 2019-10-18 |
 |
| 5.2.13 |
Single sterile barrier system with protective packaging inside |
Indicates a single sterile barrier system with protective packaging inside |
This symbol shall be placed adjacent to or in combination with symbol 5.2.1, 5.2.2, 5.2.3, 5.2.4, 5.2.5, 5.2.9 or 5.2.10. |
NOTE 1 The protective packaging located inside the sterile barrier system is designed to prevent damage to the contents or to help with aseptic presentation. It does not provide a microbial barrier to maintain sterility. NOTE 2 Additional information on sterile barrier systems can be found in ISO 11607-1[5] and ISO 11607-2[6]. |
— |
ISO 7000-3708 2019-10-18 |
 |
| 5.2.14 |
Single sterile barrier system with protective packaging outside |
Indicates a single sterile barrier system with protective packaging outside |
This symbol shall be placed adjacent to or in combination with symbol 5.2.1, 5.2.2, 5.2.3, 5.2.4, 5.2.5, 5.2.9 or 5.2.10. |
NOTE 1 The protective packaging located outside the sterile barrier system is designed to prevent damage to the sterile barrier system and the contents. The protection can be against physical hazards, particulate contamination or other environmental hazards, but it does not include a microbial barrier. NOTE 2 Additional information on sterile barrier systems can be found in ISO 11607-1[5] and ISO 11607-2[6]. |
— |
ISO 7000-3709 2019-10-18 |
 |
| 5.3 Storage |
| 5.3.1 |
Fragile, handle with care |
Indicates a medical device that can be broken or damaged if not handled carefully |
— |
— |
— |
ISO 7000-0621 2014-06-04 |
 |
| 5.3.2 |
Keep away from sunlight |
Indicates a medical device that needs protection from light sources |
— |
NOTE This symbol can also mean “Keep away from heat”. |
— |
ISO 7000-0624 2014-06-04 |
 |
| 5.3.3 |
Protect from heat and radioactive sources |
Indicates a medical device that needs protection from heat and radioactive sources |
— |
NOTE 1 This symbol can also mean “Keep away from sunlight and radioactive sources”. NOTE 2 Radioactive sources include ionizing radiation. |
— |
ISO 7000-0615 2004-01-15 |
 |
| 5.3.4 |
Keep dry |
Indicates a medical device that needs to be protected from moisture |
— |
NOTE This symbol can also mean “Keep away from rain” as referenced in ISO 7000. |
— |
ISO 7000-0626 2014-06-04 |
 |
| 5.3.5 |
Lower limit of temperature |
Indicates the lower limit of temperature to which the medical device can be safely exposed |
The lower limit of temperature shall be indicated adjacent to the lower horizontal line. |
— |
— |
ISO 7000-0534 2004-01-15 |
 |
| 5.3.6 |
Upper limit of temperature |
Indicates the upper limit of temperature to which the medical device can be safely exposed |
The upper limit of temperature shall be indicated adjacent to the upper horizontal line. |
— |
— |
ISO 7000-0533 2004-01-15 |
 |
| 5.3.7 |
Temperature limit |
Indicates the temperature limits to which the medical device can be safely exposed |
The upper and lower limits of temperature shall be indicated adjacent to the upper and lower horizontal lines. |
— |
— |
ISO 7000-0632 2014-06-04 |
 |
| 5.3.8 |
Humidity limitation |
Indicates the range of humidity to which the medical device can be safely exposed |
The humidity limitation shall be indicated adjacent to the upper and lower horizontal lines. |
— |
— |
ISO 7000-2620 2004-01-15 |
 |
| 5.3.9 |
Atmospheric pressure limitation |
Indicates the range of atmospheric pressure to which the medical device can be safely exposed |
The atmospheric pressure limitations shall be indicated adjacent to the upper and lower horizontal lines. |
— |
— |
ISO 7000-2621 2004-01-15 |
 |
| 5.4 Safe use |
| 5.4.1 |
Biological risks |
Indicates that there are potential biological risks associated with the medical device |
— |
NOTE This symbol is not to be confused with the safety sign “Biological hazard” ISO 7010-W009. |
— |
ISO 7000-0659 2004-01-15 |
 |
| 5.4.2 |
Do not re-use |
Indicates a medical device that is intended for one single use only |
— |
NOTE Synonyms for “Do not re-use” are “single use” and “use only once”. |
— |
ISO 7000-1051 2004-01-15 |
 |
| 5.4.3 |
Consult instructions for use or consult electronic instructions for use |
Indicates the need for the user to consult the instructions for use |
— |
NOTE 1 Synonym for “Consult instructions for use” is “Consult operating instructions”. NOTE 2 See also ISO 20417[15] and the safety sign ISO 7010-M002. NOTE 3 See A.16 for examples and for use in directing users to consult the electronic instructions for use. |
— |
ISO 7000-1641 2004-01-15 |
 |
| 5.4.4 |
Caution |
Indicates that caution is necessary when operating the device or control close to where the symbol is placed, or that the current situation needs operator awareness or operator action in order to avoid undesirable consequences |
The symbol variant ISO 7000-0434B (“Caution”) may be used. |
— |
“This symbol shall not be used solely to mean “consult instructions for use”. |
ISO 7000-0434A 2004-01-15 |
 |
| 5.4.5 |
Contains or presence of natural rubber latex |
Indicates the presence of dry natural rubber or natural rubber latex as a material of construction within the medical device or the packaging of a medical device |
— |
NOTE This symbol is intended to warn those people who can have allergic reactions to certain latex proteins. |
This symbol should not be used for medical devices containing “synthetic rubber”. |
Application of ISO 7000, symbol 2725 2005-09-08 |
 |
| 5.4.6 |
Contains human blood or plasma derivatives |
Indicates a medical device that contains or incorporates human blood or plasma derivatives |
The embedded cross may be deleted or replaced with another element appropriate with cultural requirements. |
— |
— |
ISO 7000-3701 2010-10-18 |
 |
| 5.4.7 |
Contains a medicinal substance |
Indicates a medical device that contains or incorporates a medicinal substance |
The embedded cross may be deleted or replaced with another element appropriate with cultural requirements. |
— |
— |
ISO 7000-3702 2019-10-18 |
 |
| 5.4.8 |
Contains biological material of animal origin |
Indicates a medical device that contains biological tissue, cells, or their derivatives, of animal origin |
— |
— |
— |
ISO 7000-3699 2019-10-18 |
 |
| 5.4.9 |
Contains biological material of human origin |
Indicates a medical device that contains biological tissue, cells, or their derivatives, of human origin |
— |
— |
— |
ISO 7000-3700 2019-10-18 |
 |
| 5.4.10 |
Contains hazardous substances |
Indicates a medical device that contains substances that can be carcinogenic, mutagenic, reprotoxic (CMR), or substances with endocrine disrupting properties |
— |
NOTE The term “substances” is used to indicate a single substance or multiple substances. |
— |
ISO 7000-3723 2019-11-01 |
 |
| 5.4.11 |
Contains nano materials |
Indicates a medical device that contains nano materials |
— |
— |
— |
ISO 7000-3703 2019-10-18 |
 |
| 5.4.12 |
Single patient multiple use |
Indicates a medical device that may be used multiple times (multiple procedures) on a single patient |
— |
— |
— |
ISO 7000-3706 2019-10-18 |
 |
| 5.5 IVD-specific |
| 5.5.1 |
In vitro diagnostic medical device |
Indicates a medical device that is intended to be used as an in vitro diagnostic medical device |
— |
NOTE For use in Europe, the full definition of “in vitro diagnostic medical device” is given in EU Regulation 2017/746.[24] Other jurisdictions can have unique definitions. |
This symbol should only be used to identify in vitro diagnostic medical devices or their accessories and not to specify that the medical device is for “in vitro use”. |
N/A |
 |
| 5.5.2 |
Control |
Indicates a control material that is intended to verify the performance of another medical device |
— |
NOTE For negative controls, use symbol 5.5.3 and for positive controls use symbol 5.5.4. |
— |
N/A |
 |
| 5.5.3 |
Negative control |
Indicates a control material that is intended to verify the results in the expected negative range |
— |
— |
— |
ISO 7000-2495 2004-01-15 |
 |
| 5.5.4 |
Positive control |
Indicates a control material that is intended to verify the results in the expected positive range |
— |
— |
— |
ISO 7000-2496 2004-01-15 |
 |
| 5.5.5 |
Contains sufficient for <n> tests |
Indicates the total number of tests that can be performed with the medical device |
The number of tests that can be performed with the medical device shall appear adjacent to the symbol. |
NOTE This symbol is suitable for use with all medical devices including in vitro diagnostic medical devices. |
— |
ISO 7000-0518 2004-01-15 |
 |
| 5.5.6 |
For IVD performance evaluation only |
Indicates an IVD medical device that is intended to be used only for evaluating its performance characteristics before it is placed on the market for medical diagnostic use |
— |
NOTE 1 A synonym is “IVD for investigational use only”. NOTE 2 A medical device that is for in vitro diagnostic performance evaluation only is not intended to be used for an in vitro diagnostic examination for medical purposes (i.e. to yield diagnostic results). |
This symbol shall not appear jointly with symbol 5.5.1. |
ISO 7000-3083 2011-10-03 |
 |
| 5.6 Transfusion/infusion |
| 5.6.1 |
Sampling site |
Indicates a medical device or blood processing application that includes a system dedicated to the collection of samples of a given substance stored in the medical device or blood container |
— |
— |
This symbol is not to be associated with a site on a patient where samples are taken. |
ISO 7000-2715 2005-09-08 |
 |
| 5.6.2 |
Fluid path |
Indicates the presence of a fluid path |
— |
NOTE The term “fluid” means a liquid or gas. |
— |
ISO 7000-2722 2005-09-08 |
 |
| 5.6.3 |
Non-pyrogenic |
Indicates a medical device that is non-pyrogenic |
— |
— |
— |
ISO 7000-2724 2005-09-08 |
 |
| 5.6.4 |
Drops per millilitre |
Indicates the number of drops per millilitre |
The number of drops per millilitre is specified; 20 is shown as an example and shall be replaced by the appropriate number of drops per millilitre. |
— |
— |
ISO 7000-2726 2005-09-08 |
 |
| 5.6.5 |
Liquid filter with pore size |
Indicates an infusion or transfusion system of the medical device that contains a filter of a particular nominal pore size |
The nominal pore size of the filter is specified; 15 is shown as an example and shall be replaced by the appropriate pore size. |
— |
— |
ISO 7000-2727 2005-09-08 |
 |
| 5.6.6 |
One-way valve |
Indicates a medical device with a valve that allows flow in only one direction |
— |
NOTE It is important for the user to know that the flow is only possible in one direction and cannot be reversed. |
— |
ISO 7000-2728 2005-09-08 |
 |
| 5.7 Others |
| 5.7.1 |
Patient number |
Indicates a unique number associated with an individual patient |
When used, the symbol shall appear adjacent to the patient number or next to a space provided to record it. |
NOTE The hash mark (#) is part of the symbol. |
— |
ISO 7000-2610 2004-01-15 |
 |
| 5.7.2 |
Patient name |
Indicates the name of the patient |
When used, the symbol shall appear adjacent to the patient name or next to a space provided to record it. |
— |
— |
ISO 7000-3726 2019-11-01 |
 |
| 5.7.3 |
Patient identification |
Indicates the identification data of the patient |
When used, the symbol shall appear adjacent to the patient identification or next to a space provided to record it. |
NOTE The question mark (?) is part of the symbol. |
— |
IEC 60417-5664 2002-10-07 |
 |
| 5.7.4 |
Patient information website |
Indicates a website where a patient can obtain additional information on the medical product |
This symbol shall be accompanied by the web address adjacent to the symbol. |
NOTE Usage is to indicate location of information available to the patient. |
— |
ISO 7000-3705 2019-10-18 |
 |
| 5.7.5 |
Health care centre or doctor |
Indicates the address of the health care centre or doctor where medical information about the patient may be found |
When used, the symbol shall appear adjacent to the address of the health care centre or doctor or next to a space provided to record it. |
NOTE The embedded cross can be deleted or replaced with another element appropriate with cultural requirements. |
— |
ISO 7001 PI PF 044 2013-05-31 |
 |
| 5.7.6 |
Date |
Indicates the date that information was entered or a medical procedure took place |
When used, the symbol shall appear adjacent to the date appropriate for the use of this symbol or next to a space provided to record it. |
— |
— |
IEC 60417-5662 2002-10-07 |
 |
| 5.7.7 |
Medical device |
Indicates the item is a medical device |
— |
NOTE For use in Europe the full definition of “medical device” is given in EU Regulation 2017/745.[23] Other jurisdictions can have unique definitions. |
— |
N/A |
 |
| 5.7.8 |
Translation |
Indicates that the original medical device information has undergone a translation which supplements or replaces the original information |
This symbol shall be accompanied by the name and address of the entity that is responsible for the translation activity adjacent to the symbol. |
NOTE If multiple symbols (i.e. Authorized Representative, Importer, Distributor, Translation, or Repackaging) identify the same responsible entity, the name and address need not be duplicated. |
This symbol shall only be used when the translation activity was undertaken by someone other than the manufacturer. |
ISO 7000-3728 2019-11-01 |
 |
| 5.7.9 |
Repackaging |
Indicates that a modification to the original medical device packaging configuration has occurred |
This symbol shall be accompanied by the name and address of the entity that is responsible for the repackaging activity adjacent to the symbol. |
NOTE 1 Depending on the authority having jurisdiction, additional information (i.e. date of repackaging) can be needed.NOTE 2 If multiple symbols (i.e. Authorized Representative, Importer, Distributor, Translation, or Repackaging) identify the same responsible entity, the name and address need not be duplicated. |
This symbol shall only be used when the repackaging activity was undertaken by someone other than the manufacturer. |
ISO 7000-3727 2019-11-01 |
 |
| 5.7.10 |
Unique device identifier |
Indicates a carrier that contains unique device identifier information |
This symbol may be used when multiple data carriers are present on the label. If used, this symbol shall be placed adjacent to the unique device identifier carrier. |
NOTE This symbol identifies the UDI carrier, including the AIDC and human readable information. |
— |
N/A |
 |