ISO / ISO 10993-1:2025 Biological evaluation of medical devices — Part 1: Requirements and general principles for the evaluation of biological safety within a risk management process

ISO 10993-1:2025 Biological evaluation of medical devices — Part 1: Requirements and general principles for the evaluation of biological safety within a risk management process

Foreword

ISO (the International Organization for Standardization) is a worldwide federation of national standards bodies (ISO member bodies). The work of preparing International Standards is normally carried out through ISO technical committees. Each member body interested in a subject for which a technical committee has been established has the right to be represented on that committee. International organizations, governmental and non-governmental, in liaison with ISO, also take part in the work. ISO collaborates closely with the International Electrotechnical Commission (IEC) on all matters of electrotechnical standardization.

The procedures used to develop this document and those intended for its further maintenance are described in the ISO/IEC Directives, Part 1. In particular, the different approval criteria needed for the different types of ISO document should be noted. This document was drafted in accordance with the editorial rules of the ISO/IEC Directives, Part 2 (see www.iso.org/directives).

ISO draws attention to the possibility that the implementation of this document may involve the use of (a) patent(s). ISO takes no position concerning the evidence, validity or applicability of any claimed patent rights in respect thereof. As of the date of publication of this document, ISO had not received notice of (a) patent(s) which may be required to implement this document. However, implementers are cautioned that this may not represent the latest information, which may be obtained from the patent database available at www.iso.org/patents. ISO shall not be held responsible for identifying any or all such patent rights.

Any trade name used in this document is information given for the convenience of users and does not constitute an endorsement.

For an explanation of the voluntary nature of standards, the meaning of ISO specific terms and expressions related to conformity assessment, as well as information about ISO's adherence to the World Trade Organization (WTO) principles in the Technical Barriers to Trade (TBT), see www.iso.org/iso/foreword.html.

This document was prepared by Technical Committee ISO/TC 194, Biological and clinical evaluation of medical devices, in collaboration with the European Committee for Standardization (CEN) Technical Committee CEN/TC 206, Biological and clinical evaluation of medical devices, in accordance with the Agreement on technical cooperation between ISO and CEN (Vienna Agreement).

This sixth edition cancels and replaces the fifth edition (ISO 10993-1:2018), which has been technically revised.

The main changes are as follows:

— this document has been completely reorganised and the title has been aligned with the risk management framework described in ISO 14971;

— content has been added to provide guidance and clarification of calculation of exposure duration;

— content has been added to provide guidance on characterization of the device and identification of biological hazards;

— the identification of biological effects (previously referred to as biological end points) has been modified;

— the term “externally communicating” has been replaced by language which reflects the specific tissue contact of device components;

— the term “effects after implantation” has been changed to “local effects after tissue contact” as some non-implanted devices also will need this type of assessment;

— Annex A has been revised to move most of the content to the main text and the remaining text in Annex A is now confined to the provision of guidance on materials characterization;

— Annex B has been added to explain the rationale for the changes to biological effects listed in Table 1 to Table 4.

A list of all parts in the ISO 10993 series can be found on the ISO website.

Any feedback or questions on this document should be directed to the user’s national standards body. A complete listing of these bodies can be found at www.iso.org/members.html.

Introduction

The aim of this document is to provide guidance and requirements for the biological evaluation of a medical device within a risk management process to protect humans from biological risks arising from the use of medical devices and the materials from which they are made. Biological risk evaluation compares the estimated biological risk against given risk criteria to determine the acceptability of the biological risk as part of the overall risk management.

Biological evaluation is primarily concerned with medical device biological safety, through consideration of risks associated with biological hazards. Nonetheless, some activities undertaken in the course of biological evaluation in addition to assessments of long-term safety can also generate information on device performance. For example, functional implant models can be used to assess long-term responses such as tissue ingrowth. Biological evaluation, as described in this document, is synonymous with biocompatibility evaluation.

Biological evaluation is conducted on the finished medical device. The principles and methods described can also be useful in the evaluation of candidate materials or prototype devices during a medical device development process, and data obtained from such evaluations can be of value in the assessment of the finished medical device.

Medical device design is wide-ranging, and, at one extreme, a medical device consists only of a single material, which can exist in more than one physical form, while at the other extreme, is a complex article consisting of numerous components made from multiple materials. Biological safety cannot be considered in isolation from the overall medical device design and can require the balancing of conflicting requirements. For example, the choice of the best material with respect to its biological safety can result in a less functional medical device.

The evaluation of biological safety is conducted in the context of the intended use of a particular medical device. Materials can be safe in one medical device and not in another. It is impossible to make generalized conclusions about the safety of a particular material for all medical applications. Biological responses that are regarded as adverse, caused by a material in one application, are not necessarily regarded as adverse in a different situation.

Physical and chemical information supports the overall biological evaluation and can be used to inform testing needs, if any. When biological testing is required, such testing is based upon in vitro, ex vivo or in vivo models. The interpretation of the results of biological tests requires caution because the inherent variability in biological responses between species and individuals means that the biological response observed in animal or cell culture models can differ from those observed in clinical use. Differences in response to the same material among individuals means that some individuals can have adverse reactions, even to well-established materials. Thus, biological evaluation is an exercise in risk management. When applied to the evaluation of candidate materials or prototype devices during a medical device development process, biological evaluation allows the informed and timely consideration of risk control measures such as use of alternative materials, manufacturing processes or designs.

The biological evaluation process described in this document draw on all available sources of information relevant to biological safety of the medical device, including post-market information. This approach allows a comprehensive review of the medical device, the identification of biological hazards and the biological harms which can arise and estimation of the associated risks. This comprehensive approach allows for the identification of any gaps in the existing data set and the consequent need for supplementary assessments (e.g. chemical analysis and hazard identification, biological testing to refine a biological risk estimate) to be conducted.

This document is supported by a wide range of test methods and guidance published in other documents in the ISO 10993 series, as well as other standards. Those who use this document can also consider more specific guidance contained in device-specific standards, where available. For some complex or novel materials or technologies, it can be difficult to use the established methods described in the ISO 10993 series. This document allows for the use of alternative procedures where scientifically justified.

The welfare of animals is very important and the selection of test methods and evolution of testing within the ISO 10993 framework is directed to continue to reduce, refine and, where possible, replace the use of animals for biological testing.

1 Scope

This document specifies the requirements and general principles governing the biological evaluation of medical devices within a risk management process according to ISO 14971.

This document applies to the biological evaluation of medical devices that have direct contact or indirect contact with either:

— a patient's body during intended use or reasonably foreseeable misuse; or

— the body of other users who are not patients, if the medical device is intended for personal protection (e.g. medical gloves, surgical masks).

Biological evaluation assesses the biological safety of the medical device by considering the biological risks associated with:

— constituents of a medical device; and

— tissue-device interactions (including physical effects).

The biological evaluation specified in this document can address the biological safety of the medical device, considering the life cycle from design and development through initial use of the finished medical device to final decommissioning or withdrawal from use. The biological evaluation considers both the biological safety of the finished device in first use, and the significance of any changes to the medical device which can occur throughout the life cycle. However, the evaluation of risks related to environmental impacts of decommissioning of medical devices are not within the scope of this document. This document does not mandate re-testing of medical devices that are already on the market and have established and acceptable safety profiles (see 6.6.2).

This document can be useful to support clinical or usability evaluations of medical devices. For example, a biological evaluation is a pre-requisite for conducting a clinical trial. This means that principles outlined in this document can be applied to the evaluation of prototype or development stage devices, as well as to finished medical devices.

Other parts of the ISO 10993 series cover specific aspects of biological evaluation, such as chemical characterization, biological testing, sample preparation, animal welfare and toxicological risk assessment.

For some types of medical devices, specific requirements from other standards (outside the ISO 10993 series) can be considered with a justification for the approach taken if there are differences between the requirements of the ISO 10993 series and those provided in other standards. For example, the ISO 18562 series provides specific requirements for biological evaluation of breathing gas pathway medical devices and ISO 7405 provides specific requirements for biological evaluation of dental devices.

The evaluation of risks related to infectious agents [e.g. bacteria, moulds, yeasts, viruses, transmissible spongiform encephalopathy (TSE) agents] is not within the scope of this document.

NOTE 1

The evaluation of bacterial endotoxins is addressed by ISO 11737-3.

NOTE 2

The evaluation of risks related to viruses, TSE agents and other pathogens originating from materials of animal origin is addressed by the ISO 22442 series.

2 Normative references

The following documents are referred to in the text in such a way that some or all of their content constitutes requirements of this document. For dated references, only the edition cited applies. For undated references, the latest edition of the referenced document (including any amendments) applies.

  • ISO 10993-2 Biological evaluation of medical devices — Part 2: Animal welfare requirements
  • ISO 10993-3 Biological evaluation of medical devices — Part 3: Tests for genotoxicity, carcinogenicity and reproductive toxicity
  • ISO 10993-4 Biological evaluation of medical devices — Part 4: Selection of tests for interactions with blood
  • ISO 10993-5 Biological evaluation of medical devices — Part 5: Tests for in vitro cytotoxicity
  • ISO 10993-6 Biological evaluation of medical devices — Part 6: Tests for local effects after implantation
  • ISO 10993-7 Biological evaluation of medical devices — Part 7: Ethylene oxide sterilization residuals
  • ISO 10993-9:2019 Biological evaluation of medical devices — Part 9: Framework for identification and quantification of potential degradation products
  • ISO 10993-10 Biological evaluation of medical devices — Part 10: Tests for skin sensitization
  • ISO 10993-11 Biological evaluation of medical devices — Part 11: Tests for systemic toxicity
  • ISO 10993-12 Biological evaluation of medical devices — Part 12: Sample preparation and reference materials
  • ISO 10993-13 Biological evaluation of medical devices — Part 13: Identification and quantification of degradation products from polymeric medical devices
  • ISO 10993-14 Biological evaluation of medical devices — Part 14: Identification and quantification of degradation products from ceramics
  • ISO 10993-15 Biological evaluation of medical devices — Part 15: Identification and quantification of degradation products from metals and alloys
  • ISO 10993-16 Biological evaluation of medical devices — Part 16: Toxicokinetic study design for degradation products and leachables
  • ISO 10993-17:2023 Biological evaluation of medical devices — Part 17: Toxicological risk assessment of medical device constituents
  • ISO 10993-18:2020 Biological evaluation of medical devices — Part 18: Chemical characterization of medical device materials within a risk management process
  • ISO/TS 10993-19 Biological evaluation of medical devices — Part 19: Physico-chemical, morphological and topographical characterization of materials
  • ISO 10993-23 Biological evaluation of medical devices — Part 23: Tests for irritation
  • ISO 14971:2019 Medical devices — Application of risk management to medical devices

3 Terms and definitions

For the purposes of this document, the following terms and definitions apply.

ISO and IEC maintain terminology databases for use in standardization at the following addresses:

— ISO Online browsing platform: available at https://www.iso.org/obp

— IEC Electropedia: available at https://www.electropedia.org/

3.1 biocompatibility

ability of a medical device (3.28) or material (3.27) to perform with an appropriate host response in a specific application

3.2 biological effect

observable or measurable response of cells or a living organism to the physical characteristics or chemical constituents (3.15) of a medical device (3.28) including a physical [e.g. nanomaterial (3.30)] or biological agent (e.g. animal tissue (3.40) derivative)

3.3 biological equivalence

circumstance where the chemical and physical characteristics and manufacturing process of a medical device (3.28) or material (3.27) under evaluation can be shown to be sufficiently similar to a comparator medical device or material for which there are existing data demonstrating biological safety (3.11) in a relevant type and duration of body contact and that no new or increased biological risks (3.8) are identified for the medical device or material under evaluation

3.4 biological evaluation

risk management (3.37) process associated with biological risks (3.8) of a medical device (3.28) or material (3.27)

3.5 biological risk evaluation

process of comparing the biological risk (3.8) of a medical device (3.28) or material (3.27) against defined criteria to determine the acceptability of the biological risk

Biological risk evaluation is a subset of the overall biological evaluation (3.4).

ISO 14971:2019, 3.23, modified — “biological” has been added before “risk” in the term and in the definition, and “biological risk of a medical device or material” has replaced “estimated risk” and “defined criteria” has replaced “given risk criteria” in the definition.

3.6 biological harm

injury to humans from one or more adverse biological effects (3.2) associated with a medical device (3.28) or material (3.27)

3.7 biological hazard

potential source of biological harm (3.6)

Biological hazards relevant to the intended use (3.23) can arise from either exposure to a constituent (3.15) of a medical device (3.28) or the physical characteristics [e.g. texture, stiffness, configuration (3.14)] of the medical device.

ISO 14971:2019, 3.4, modified — “biological” has been added before “hazard” in the term and before “harm” in the definition, and NOTE 1 to entry has been added.

3.8 biological risk

combination of the probability of occurrence of biological harm (3.6) and the severity of that biological harm

ISO 14971:2019, 3.18, modified — “biological” has been added before “risk” in the term and before “harm” in the definition.

3.9 biological risk analysis

systematic use of available information to identify biological hazards (3.7) and to estimate biological risks (3.8)

ISO 14971:2019, 3.19, modified — “biological” has been added before “risk” in the term, and before “hazards” and “risks” in the definition.

3.10 biological risk assessment

overall process comprising a biological risk analysis (3.9) and a biological risk evaluation (3.5)

ISO 14971:2019, 3.20, modified — “biological” has been added before “risk” in the term, and before “risk assessment” and “risk evaluation” in the definition.

3.11 biological safety

freedom from unacceptable biological risk (3.8) in the context of the intended use (3.23) or reasonably foreseeable misuse (3.34)

ISO 14971:2019, 3.26, modified — “biological” has been added before “safety” in the term and before “risk” in the definition, and “in the context of the intended use (3.23) or reasonably foreseeable misuse” has been added to the definition.

3.12 biologically hazardous situation

circumstances which result in exposure to one or more biological hazards (3.7)

3.13 chemical characterization

process of obtaining chemical information, accomplished either by information gathering or by information generation, for example, by literature review or chemical testing

ISO 10993-18:2020, 3.6

3.14 configuration

geometry, shape, size and relative arrangement of the parts of the medical device (3.28)

3.15 constituent

chemical that is present in or on the finished medical device (3.20) or its materials (3.27) of construction

Constituents can be intentionally or unintentionally added chemicals or compounds, such as: additives (e.g. plasticizers, lubricants, stabilizers, anti-oxidants, colouring agents, fillers), manufacturing process residues (e.g. monomers, catalysts, solvents, sterilant and cleaning agents), degradation (3.17) products, reaction products, or impurities or contaminants.

ISO 10993-17:2023, 3.4, modified — “reaction products” has been added to and “(e.g. byproducts or side products)” has been removed from Note 1 to entry.

3.16 contact day

day in which a medical device (3.28) or component thereof comes into contact with tissue (3.40) including circulating blood, irrespective of the length of time of that contact within the day

3.17 degradation

breakdown of a material (3.27) by physical or chemical means

3.18 direct contact

physical contact of a medical device (3.28) or component thereof, with tissue (3.40) including circulating blood

3.19 extractable

substance (3.39) that is released from a medical device (3.28) or material (3.27) when the medical device or material is extracted using laboratory extraction conditions and vehicles

ISO 10993-18:2020, 3.16, modified — ‘of construction’ has been removed from the definition.

3.20 finished medical device

medical device (3.28) that is in the state in which it is intended to be used, following all manufacturing processes such as packaging, storage and sterilisation

3.21 implant

medical device (3.28) or component thereof, that is intended to be introduced into the human body or to replace an epithelial surface or the surface of the eye, by means of surgical intervention and that is intended to remain in place after the procedure

The duration of an implant remaining in the body is dependent on the clinical need.

3.22 indirect contact

medical device (3.28) or component thereof, having no physical contact with tissue (3.40), through which a fluid (liquid or gas), semi-solid or solid passes prior to contacting tissue

3.23 intended use

use for which a medical device (3.28) is intended according to the specifications, instructions and information provided by the manufacturer (3.26)

The intended medical indication, patient population, part of the body or type of tissue (3.40) interacted with, user (3.43) profile, use environment, and operating principle are typical elements of the intended use.

ISO 14971:2019, 3.6, modified — “product, process or service” has been replaced with “medical device” in the definition.

3.24 leachable

substance (3.39) that is released from a medical device (3.28) or material (3.27) during its clinical use

ISO 10993-18:2020, 3.22, modified — Note 1 to entry has been deleted.

3.25 life cycle

series of all phases in the life of a medical device (3.28), from the initial conception to final decommissioning and disposal

Terms for phases within the life cycle can include shelf life, service life, use life, lifetime or similar terms.

ISO 14971:2019, 3.8, modified — Note 1 to entry has been added.

3.26 manufacturer

natural or legal person with responsibility for the design or manufacture of a medical device (3.28) with the intention of making the medical device available for use, under their name, whether or not such a medical device is designed or manufactured by that person themselves or on their behalf by another person(s)

ISO 14971:2019, 3.9, modified — “his” has been changed to “their” and “himself” has been changed to “themselves” in the definition, and Notes 1, 2, 3, 4, 5, 6 and 7 to entry have been deleted.

3.27 material

synthetic or natural substance (3.39), including polymer, metal or alloy, ceramic, composite, tissue (3.40), used as a medical device (3.28) or any part thereof

3.28 medical device

instrument, apparatus, implement, machine, appliance, implant (3.21), reagent for in vitro use, software, material (3.27) or other similar or related article, intended by the manufacturer (3.26) to be used, alone or in combination, for human beings, for one or more of the specific medical purpose(s) of

— diagnosis, prevention, monitoring, treatment or alleviation of disease,

— diagnosis, monitoring, treatment, alleviation of or compensation for an injury,

— investigation, replacement, modification, or support of the anatomy or of a physiological process,

— supporting or sustaining life,

— control of conception,

— disinfection of a medical device (3.28),

— providing information by means of in vitro examination of specimens derived from the human body,

and which does not achieve its primary intended action by pharmacological, immunological or metabolic means, in or on the human body, but which can be assisted in its intended function by such means

Products which can be considered to be medical device in some jurisdictions but not in others include:

— disinfection substances (3.39);

— aids for persons with disabilities;

— devices incorporating animal or human tissues (3.40);

— devices for in vitro fertilization or assisted reproduction technologies.

ISO/IEC Guide 63:2019, 3.7, modified — “and/or” has been replace with "“or” in Note 1 to entry.

3.29 medical device family

group of medical devices (3.28) manufactured by or for the same organization and having the same basic design, materials (3.27), manufacturing process and performance characteristics related to safety, intended use (3.23) and function

ISO 13485:2016, 3.12, modified — “materials and manufacturing process” has been added to the definition.

3.30 nanomaterial

material (3.27) with any external dimension in the nanoscale or having internal structure or surface structure in the nanoscale, where nanoscale is in the range of approximately 1 nm to 100 nm

ISO/TR 10993-22:2017, 3.7, modified — “where nanoscale is in the range of approximately 1 nm to 100 nm” has been added to the definition and Notes 1, 2 and 3 to entry have been deleted.

3.31 non-contacting

indicates that a medical device (3.28) or component thereof, has neither direct contact (3.18) nor indirect contact (3.22) with tissue (3.40)

3.32 novel material

material (3.27) that has not previously been used in a medical device (3.28) for which biological safety (3.11) has been established relevant to the intended use (3.23)

3.33 physical and chemical information

knowledge regarding a finished medical device (3.20) or material (3.27) including the geometric, chemical and physical properties and the identity of the supplier of a material or component

3.34 reasonably foreseeable misuse

use of a medical device (3.28) in a way not intended by the manufacturer (3.26), but which can result from readily predictable human behaviour

Readily predictable human behaviour includes the behaviour of all types of users (3.43), e.g. lay and professional users.

Reasonably foreseeable misuse can be intentional or unintentional.

ISO 14971:2019, 3.15, modified — “product or system” has been replaced with “medical device in the definition”.

3.35 residual biological risk

biological risk (3.8) remaining after risk control measures have been implemented

ISO 14971:2019, 3.17, modified — “biological” has been added before “risk” in the term and in the definition.

3.36 risk estimation

process used to assign values to the probability of occurrence of harm and the severity of that harm

ISO 14971:2019, 3.22

3.37 risk management

systematic application of management policies, procedures and practices to the tasks of analysing, evaluating, controlling and monitoring biological risk (3.8)

ISO 14971:2019, 3.24, modified — “biological” has been added before “risk” in the definition.

3.38 sensitive population

persons who are more susceptible to biological harm (3.6) due to exposure to hazardous substances (3.39) than the average, healthy person

These populations can include persons who are: young infants, very elderly, chronically ill, immunocompromised, pregnant women or women of childbearing age.

3.39 substance

single chemical element or compound, or a complex structure of compounds

ISO 10993-9:2019, 3.5

3.40 tissue

organization of cells or extra-cellular components

This can include epithelial tissue, muscle tissue, nervous tissue and connective tissue (including adipose, blood, bone, cartilage, and lymphatic tissue).

ISO 22442-1:2020, 3.9, modified — “and/or” has been changed to “or”, “constituents” has been changed to “components” in the definition and Note 1 to entry has been added.

3.41 total exposure period

number of contact days (3.16) from the first to the last use of a medical device (3.28)

3.42 toxicity

capability of a constituent (3.15) to cause a biological harm (3.6) under specified conditions of exposure or dose

3.43 user

person who comes into contact [including indirect contact (3.22)] with a medical device (3.28) during its use, including patients or persons who are required to operate or handle the medical device during its use

4 General principles

4.1 Biological evaluation within the ISO 14971 risk management framework

The biological evaluation (biocompatibility evaluation) of a medical device shall be conducted within a risk management process in accordance with ISO 14971 (see Figure 1), which involves the identification of biological hazards, the estimation of the associated biological risks and the determination of their acceptability (see Clause 6 and Clause 7).

The biological evaluation shall be planned (see Clause 5), conducted and reported by competent personnel, capable of making informed decisions based upon scientific data and a knowledge of relevant medical devices. Competence shall be demonstrated and documented on the basis of education, training, skills and experience (see Clause 9).

NOTE 1

Biological evaluation tasks can be performed by representatives of several functions or organizations, each contributing their specialist knowledge and competence.

To understand the relationship between ISO 14971 and the ISO 10993 series, it is useful to apply the risk management terminology of ISO 14971 as follows:

— Biological hazards can arise from the presence of constituents which are in or on the medical device, or from physical characteristics of the medical device.

— A biologically hazardous situation arises when direct contact or indirect contact with a medical device causes exposure to a hazardous constituent or physical characteristic.

— The biologically hazardous situation can lead to a biological harm.

— The biological risk is the combination of the probability of occurrence and severity of the adverse biological effect resulting from exposure to the hazardous constituent or physical characteristic.

The following summarizes the application of the ISO 14971 risk management principles to the evaluation of biological safety.

Figure 1
ISO 14971 risk management framework

Biological risk analysis requires:

— description and categorization of the medical device including its configuration, composition, physical characteristics and body contact during use (including consideration of reasonably foreseeable misuse) (see 6.4);

— identification of biological hazards, biologically hazardous situations and possible biological harms arising from physical and chemical information (due to the potential availability of toxic constituents or hazardous physical characteristics) (see 6.3) and the use of the medical device (see 6.4.1);

— identification of relevant biological effects (see 6.3, 6.4 and 6.5).

Biological risk estimation requires consideration of likely exposure to hazardous constituents or effects of physical characteristics of the medical device. If existing data are insufficient, further physical, chemical or biological testing (see 6.8) can be required to estimate biological risks (see 6.9).

Biological risk evaluation is the determination of the acceptability of the estimated biological risks (see Clause 7).

Biological risk control (see Clause 8) consists of any actions taken to reduce identified biological risks. Risk control measures shall be applied in the following priority order:

  1. a) inherently safe design and manufacturing;
  2. b) protective measures in the medical device itself or in the manufacturing process;
  3. c) information for safety and, where appropriate, training to users.

Re-analysis of biological risks (which can include additional testing) is necessary to confirm that the identified biological risk was reduced, and no unacceptable new risks have been introduced by implementation of biological risk control measures (see Clause 8).

The evaluation of overall residual risk is conducted following the implementation of all risk controls (see ISO 14971:2019).

NOTE 2

Biological safety is determined based on a biological evaluation of all of the available data rather than the findings for any individual biological effect.

A biological evaluation report shall document the biological evaluation process, including the outcome and conclusions on the biological safety of the finished medical device (see Clause 8).

Production and post-production activities include monitoring of post-market feedback from users or patients and evaluation of any adverse biological effects observed. This information can be used to update and revise the biological evaluation (see 6.9) and, if required, provide input to biological risk control e.g. design changes (see Clause 8).

4.2 Biological evaluation process

The overall process for the conduct of a biological evaluation is depicted in Figure 2. The biological evaluation process shall be established, implemented, documented and maintained over the life cycle of a medical device (see 4.3 and Clause 10).

NOTE

The flowchart in Figure 2 is provided as a visual aid but does not capture all necessary considerations. Refer to the identified clauses and subclauses of this document when using this flowchart.

Figure 2
Process for conducting a biological evaluation

4.3 Medical device life cycle

The biological evaluation of a medical device shall consider its life cycle.

In the selection of materials to be used in the manufacture of a medical device, the first consideration shall be fitness for purpose with regard to characteristics and properties of the material, which include chemical, toxicological, physical, electrical, morphological and mechanical properties. Further guidance on material selection can be found in Clause A.2.

The potential for change in the characteristics of the medical device during transport, storage or use (e.g. where there is anticipated degradation or potential for physical failures causing exposure to previously unevaluated materials or constituents) shall be considered.

For reusable devices, evaluation of biological safety shall take into consideration the number of processing cycles based on the expected use life of the device. If the maximum number of processing cycles is not specified by the manufacturer, the biological risk control measures shall be justified and documented.

Post-market data can show that the biological safety of the finished medical device can be different from that anticipated during its development. Therefore, biological evaluations shall be reviewed and updated as required by post-market information or where new or changed biological risks are identified. This is part of the overall risk management process.

4.4 Animal welfare

Any in vivo testing shall be conducted in accordance with ISO 10993-2. The conduct of any in vivo biological testing shall be minimized to that which is strictly necessary to complete the biological evaluation. In the following order, where possible, the biological evaluation shall:

— refine biological risk estimation by considering additional information in the literature, conducting additional data analysis or generating additional chemical and physical information (see 6.8.2);

— replace the use of animal testing with alternative techniques (e.g. in vitro or in silico) where these methods provide equally relevant information to that obtained from in vivo models;

— reduce the number of animals used to a minimum to obtain information from fewer animals or to maximize the amount of data collected from each animal in a way that is compatible with the study objectives; and

— refine the way experiments are carried out to minimize animal pain and distress.

The approach taken shall be justified and documented.

In vivo testing shall not be carried out where:

  1. a) valid results are available from relevant non-clinical studies that have been carried out previously on the medical device or on a medical device where biological equivalence has been established (see 6.7);
  2. b) valid results relevant to the specific biological effects under evaluation are available from the existing clinical data on the medical device or on a medical device where biological equivalence has been established (see 6.7); in assessing the data, both the quality of the data and the type of available clinical information should be taken into account; or
  3. c) risk evaluation indicates an unacceptable risk to the animal based on the existing information (e.g. expected skin corrosion based on pH).

NOTE

Additional information on validity of testing is included in ISO 10993-2, ISO 10993-11 and ISO 14155.

5 Biological evaluation plan

The biological evaluation shall be planned and the plan documented (see also ISO 14971). The biological evaluation plan should include or refer to:

  1. a) a description of the finished medical device or medical device family under evaluation, including materials of construction, configuration, manufacturing processes and other processes to which the medical device is exposed throughout its life cycle (e.g. sterilization, reprocessing);
  2. b) intended use;
  3. c) reasonably foreseeable misuse;
  4. d) medical device categorization (including frequency, duration, and type of body contact) (see 6.4);
  5. e) identification of characteristics related to biological safety (see 6.2);
  6. f) identification of biological hazards, biologically hazardous situations and potential biological harms (see 6.3 and 6.5);
  7. g) criteria for acceptability of specific biological risks associated with the medical device;

    NOTE 1

    See ISO 14971 and ISO/TR 24971 for general guidance on the development of criteria for risk acceptability considering relevant parts of the ISO 10993 series.

    EXAMPLE

    Acceptability criteria can be expressed in terms of:

    — the absence of specific biological hazards;

    — the combination of maximum tolerable severity and probability of occurrence of relevant biological harms;

    — the expected outcome from a conducted test.

  8. h) gap analysis of relevant existing data (see 6.6);
  9. i) the strategy and planned content for the biological evaluation report of the medical device, including any requirements for additional data generation;
  10. j) criteria to evaluate the overall biological safety conclusions for the medical device;
  11. k) aspects of the biological evaluation requiring specific technical competencies (e.g. analytical testing, biological testing, toxicology, engineering); and
  12. l) the personnel responsible for the biological evaluation.

NOTE 2

A plan can be an evolving document, which starts as an outline and is refined as information is gathered (see 6.2, 6.3 and 6.6).

6 Biological risk analysis

6.1 General approach

The biological evaluation shall address the medical device life cycle from design and development through initial use to final decommissioning or withdrawal from use (see 4.3). This evaluation shall take all available relevant information into account, including intended use, reasonably foreseeable misuse, physical and chemical information, preclinical tests, clinical investigations, post-market experience, and any other relevant information.

Consideration of sources of data should be broadly based. Data can be drawn from raw material supplier information, information held by the manufacturer such as customer feedback and test reports, or from publicly available information such as adverse event databases, toxicology databases or scientific and clinical literature. Information can include data on relevant medical devices. Where sources of data are incomplete or less informative, the quality and relevance of the data should be justified and documented.

NOTE 1

The activities described in 6.2 to 6.5 can be conducted in parallel.

NOTE 2

The scope of the biological risk analysis can be broad (as for the development of a new medical device with which a manufacturer has little or no experience) or the scope can be limited (as for analysing the impact of a change to an existing medical device for which information already exists in the manufacturer’s files).

6.2 Identification of characteristics related to biological safety

The information needed for the biological evaluation is principally determined by the type, frequency and duration of contact (see 6.4) and the biological hazards identified for the medical device or material (see 6.5).

When identifying characteristics related to biological safety, consideration shall also be given to the following:

  1. a) the device design and materials of construction (i.e. all materials with direct contact or indirect contact) and their interactions;
  2. b) physical characteristics, including but not limited to, porosity, particle size, shape and surface morphology (see ISO/TS 10993-19);

    NOTE 1

    Nanomaterials can pose specific challenges to the biological evaluation of a medical device due to their potentially unique properties (see ISO/TR 10993-22).

  3. c) impact of manufacturing processes on materials of construction (e.g. passivation, laser marking, sterilization);
  4. d) introduction of additional constituents during manufacturing and reprocessing activities if applicable (e.g. additives, process contaminants or sterilization residues);

    NOTE 2

    ISO 10993-7 includes requirements for ethylene oxide sterilization residuals.

  5. e) primary packaging materials, which can transfer constituents to the medical device;

    NOTE 3

    In some cases, constituents from other packaging components (e.g. glues, adhesives) can be introduced through the primary packaging and impact biological risk;

  6. f) interactions with other medical devices or medicines which contact the device, which can transfer constituents to or impact the physical characteristics of tissue-contacting components of the medical device;
    EXAMPLE

    Parenteral container with liquids contained within, ultrasound transducer with lubricating gel, bearing surface interactions, cardiac leads interaction with pacemaker, prosthetic heart valve loading tools, crimping or refolding tools.

  7. g) constituents which are available on the device surface or as constituents released from the finished medical device (see ISO 10993-18 and ISO 10993-17);
  8. h) where applicable, potential degradation products (see ISO 10993-9 for general principles, and ISO 10993-13, ISO 10993-14 and ISO 10993-15 for degradation products from polymers, ceramics, and metals and alloys, respectively); and

    NOTE 4

    Degradation can be intentional or unintentional.

  9. i) consequences of changes in the performance and characteristics of the finished medical device over its life cycle (see 4.3 and Clause 10).

Further information on material characterization can be found in Annex A, ISO 10993-18 and ISO/TS 10993-19.

NOTE 5

ISO/TR 24971:2020 includes further information on identifying hazards and characteristics related to safety.

6.3 Identification of biological hazards, biologically hazardous situations and potential biological harms

Gathering physical and chemical information on the medical device or component and on the materials used in the manufacturing processing (including sterilization) is a crucial first step in the biological evaluation.

The extent of physical and chemical information required depends on the type and duration of contact and the biological hazards presented. At a minimum, the chemical characterization shall address the materials of construction and processing aids or additives used in manufacturing. In addition, it is possible that some physical characterization information will be needed, especially for implanted medical devices or medical devices in contact with tissues including blood. At a minimum, physical characterization shall address general surface characteristics and potential hazardous attributes if they can induce biological risk.

This information allows identification of biological hazards in the form of the presence of constituents at potentially toxic levels or hazardous physical characteristics of the device.

All known and foreseeable biological hazards shall be identified from the medical device characterization data. This involves:

— identifying hazardous constituent(s) from available information and estimating the quantity present in the finished medical device, and

— identifying any hazardous physical characteristics and associated adverse biological response (biological harm) which the physical characteristic can cause.

If medical device characterization data indicates the presence of a known hazardous constituent (e.g. sensitizer, carcinogen or neurotoxin) then the risk shall be evaluated in addition to the other biological effects appropriate to the device categorization (as indicated in Table 1, Table 2, Table 3 and Table 4).

Biological hazards associated with physical attributes can be more difficult to quantify and characterize than constituents (where an estimate of the available mass provides quantification). Possible factors to consider include the number and size of particulates released into the tissue and metrics such as porosity, texture and overall surface area of the physical attribute, as well as the specific tissue contact.

For the identified biologically hazardous situations, the use of quantitative information allows a quantitative estimate of biological risks (see 6.9), however, qualitative information can be adequate for biological risk assessment.

Chemical characterization shall be conducted in accordance with ISO 10993-18. Relevant physical attributes shall be characterized. ISO/TS 10993-19 provides information on physical characterization.

Where the potential for degradation exists under the conditions of manufacture, sterilization, transport, storage, reprocessing or intended use of the medical device, the presence and characteristics of degradation products shall be assessed, as applicable, to identify biological hazards associated with degradation (see 6.5.11.6).

The presence of a biological hazard does not necessarily lead to a biological harm. Determination of biological risk requires assessment of whether a biologically hazardous situation exists.

EXAMPLE 1

Biologically hazardous situations can include:

— the presence of released constituents which are known to bioaccumulate;

— the presence of released harmful additives (e.g. some plasticizers, pigments), catalysts and other constituents (e.g. unreacted monomers);

— the presence of hazardous particles released from materials;

— the presence of potentially toxic amounts of cobalt (into surrounding tissue and systemic distribution) due to wear (e.g. third body wear) of cobalt alloy joint implants.

EXAMPLE 2

Examples that do not lead to biologically hazardous situations can include:

— the presence of harmful additives, catalysts and other constituents or hazardous particles which are contained within non-degradable materials, and remain fully sequestered in the matrix material during use;

— the presence of toxic components inside a medical device (e.g. pacemaker battery), proven to be hermitically sealed inside the medical device and physically separated from the surrounding tissue and circulatory system.

6.4 Categorization of medical device and determination of scope of evaluation

6.4.1 General

The intended use of the medical device under evaluation shall be documented. The intended use should include the intended medical indication, patient population, part of the body or type of tissue interacted with, use environment and operating principle. This information shall be used to categorize the medical device in terms of the type and duration of body contact. The categorization enables the identification of potential biological effects and the selection of appropriate data sets for biological risk estimation and evaluation of the biological response.

Categorization of the medical device shall also consider the potential impact of reasonably foreseeable misuse in accordance with ISO 14971, where there is information available (e.g. post-market surveillance data, clinical literature) to suggest that the device can be used outside of its intended use, and this misuse is identified as being systematic.

EXAMPLE 1

Reasonably foreseeable misuse examples include:

— the use for longer than the period intended by the manufacturer, resulting in a longer duration of exposure (see 6.4.2);

— the use at a different body contact site than intended by the manufacturer (see 6.4.4);

— the use in a different population than intended by the manufacturer.

Body contact can be either direct contact or indirect contact. Direct contact can occur with physical contact in or on the body (e.g. a hip implant or wound dressing), or outside the body (e.g. a blood oxygenating circuit, devices intended for temporary storage and transport of organs for transplantation). Indirect contact can occur by means of transport of constituents within a fluid (gas or liquid) or semi-solid which passes from the medical device to the tissues (e.g. infusion equipment, ventilators).

NOTE

The ISO 18562 series addresses the biological evaluation for gas pathway devices or components.

Some medical devices, such as those with multiple intended uses, can fall into more than one body contact or duration category. In those instances, evaluation appropriate to all applicable categories shall be carried out or the worst-case category shall be justified.

For medical devices that include components that fall into different categories (e.g. both a long-term duration component, and a limited duration delivery system that is used only during a surgical procedure to place the long-term component), either the components shall be categorised separately or, alternatively, the worst-case category shall be applied to all components if justified.

EXAMPLE 2

A delivery system that is used only during a surgical procedure to place the implantable medical device is assessed separately from the implantable medical device.

6.4.2 Exposure duration categories

Medical devices shall be categorized in accordance with the anticipated duration of direct contact or indirect contact. With multiple exposures to a device, the duration category shall be based on the total exposure period on a single patient.

If a material or medical device can be placed in more than one duration category, requirements applicable to the longest contact duration shall apply.

Duration categories shall be determined as follows:

  1. a) limited exposure: a medical device that has a total exposure period of less than or equal to 24 h;

    NOTE 1

    Further information on evaluation of medical devices in very brief duration tissue contact is provided in 6.5.11.2.

  2. b) prolonged exposure: a medical device that has a total exposure period of more than 24 h but not more than 30 d; and
  3. c) long-term exposure: a medical device that has a total exposure period of more than 30 d.

NOTE 2

These time periods are for categorization only, and other durations can be justified for risk assessment (e.g. minutes or hours of use) or testing (e.g. hours of use) if applicable, as described in relevant standards, such as ISO 10993-5, ISO 10993-17 and ISO 10993-18.

The rationale for the approach taken for determination of the total exposure period shall be justified and documented. Where device-specific standards provide guidance for categorization, they should be considered.

NOTE 3

For certain medical devices where constituents can remain after removal of the medical device, recategorization following gap analysis (see 6.6) to consider longer exposure can be needed such as for absorbent creams (see 6.5.11.4) or delivery systems with lubricants or coatings.

6.4.3 Calculation of exposure duration for categorization of medical devices

6.4.3.1 Daily contact

Daily contact is when a medical device contacts the body every day for any portion of a day. For these medical devices, the total exposure period is the number of calendar days from the first to the last use of the medical device or a replacement of the same medical device on a single patient.

EXAMPLE 1

A continuous positive airway pressure (CPAP) machine which is used every day for five hours per day for a lifetime is a long-term exposure medical device.

EXAMPLE 2

A single use disposable feeding tube set used with a feeding pump system is replaced multiple times in a single patient over the period of treatment (e.g. two hours of use every day for 10 days would be categorized as prolonged or two hours of use every day for more than 30 days would be categorized as long-term).

6.4.3.2 Intermittent contact

Intermittent contact is the use of the device where there is a period of at least 24 hours between consecutive tissue contacts. This type of contact can be for the repeated use of one medical device or a replacement of the same medical device under consideration.

For these medical devices, the total exposure period is calculated by adding the number of contact days from the first to the last use of the medical device on a single patient. This calculation of total exposure duration applies regardless of how long a single device is used for on a particular day.

EXAMPLE 1

Medical devices used a few times a year (e.g. one hour treatment period, every six months for a maximum of three years), would be considered six days and would be categorized as prolonged unless otherwise justified.

EXAMPLE 2

Medical devices used for repeated treatment sessions (e.g. one hour treatment period, once every three weeks for a total of up to 10 sessions), would be considered 10 d and would be categorized as prolonged.

6.4.3.3 Bioaccumulation

The potential for bioaccumulation of medical device constituents shall be considered for all duration categories. Where bioaccumulation is expected (e.g. a raw material contains perfluoro-heptanoic acid, a substance known to bioaccumulate in humans), the contact duration of the medical device shall be long-term unless otherwise justified.

During the subsequent biological evaluation of the medical device, where risk for bioaccumulation is identified, the contact duration of the medical device shall be long-term unless otherwise justified.

6.4.4 Body contact and biological effects for consideration

6.4.4.1 General

Medical devices shall be categorized in accordance with their type and duration of body contact. The associated biological effects are outlined in 6.4.4.2 to 6.4.4.6.

6.4.4.2 Non-contacting medical devices

Non-contacting medical devices are medical devices (or components) which have neither direct contact nor indirect contact with the body. Absence of body contact means that no biological evaluation is required.

EXAMPLE

Diagnostic software, X-ray generators used for medical imaging, in vitro diagnostic (IVD) medical devices (assays, reagents and analysers) used to conduct measurements on tissue samples remote from the patient.

NOTE 1

Medical devices used to collect a tissue sample, such as needles, scalpels or lancets are considered to be body contacting.

NOTE 2

Medical devices used to store a tissue sample or body fluid, that is intended to have subsequent body contact, such as organ perfusion systems, are considered to have indirect contact with the body.

6.4.4.3 Medical devices in contact with intact skin

Medical devices in contact with intact skin are medical devices which have only direct contact or indirect contact with intact, uncompromised skin.

EXAMPLE

Electrodes, external prostheses, fixation tapes and compression bandages.

For these medical devices, the biological effects listed in Table 1 shall be evaluated.

Table 1
Biological effects for medical devices in contact with intact skin
Duration Biological effecta b c
Cytotoxicity (6.5.2) Sensitization (6.5.3) Irritation (6.5.4) Systemic
toxicity (6.5.5)d
Local effects after tissue contact (6.5.6) Genotoxicity (6.5.7) Carcinogenicity (6.5.8) Haemocompatibility (6.5.9)
A – limited (≤24 h) E E E          
B – prolonged (>24 h to 30 d) E E E          
C – long-term (>30 d) E E E          

Key

E biological effects to be evaluated

a It is important to emphasize that identification of an effect for evaluation requires consideration of existing data sets first and only if these are insufficient for biological risk assessment additional testing should be considered. This table is not a checklist for testing (see 6.5.1 and 6.6).

b For some medical devices it can be appropriate to evaluate additional biological effects other than those indicated for evaluation in this table, for example medical devices manufactured from a novel material or intended for use in certain sensitive populations (e.g. young infants, pregnant women, immunocompromised individuals).

c It is also important to consider if there are specific circumstances which require evaluation of other biological effects (see 6.5.10) or consideration of other factors (see 6.5.11).

d Evaluations for systemic toxicity should reflect the duration of use (see 6.5.5).

6.4.4.4 Medical devices in contact with intact mucosal membrane

Medical devices in contact with intact mucosal membrane are surface-contacting medical devices having direct contact or indirect contact with intact mucosal membranes.

EXAMPLE

Contact lenses, urinary catheters, intravaginal and intra-intestinal devices (stomach tubes, sigmoidoscopes, colonoscopes, gastroscopes), endoscopic irrigation systems and tubing, endotracheal tubes, bronchoscopes, some dental prostheses and orthodontic devices.

For these medical devices, the biological effects listed in Table 2 shall be evaluated.

Table 2
Biological effects for medical devices in contact with intact mucosal membranes
Duration Biological effecta b c
Cytotoxicity (6.5.2) Sensitization (6.5.3) Irritation (6.5.4) Systemic toxicity (6.5.5)e Local effects after tissue contact (6.5.6)d Genotoxicity (6.5.7) Carcinogenicity (6.5.8) Haemocompatibility (6.5.9)
A – limited (≤24 h) E E E          
B – prolonged (>24 h to 30 d) E E E E E E    
C – long-term (>30 d) E E E E E E E  

Key

E biological effects to be evaluated

a It is important to emphasize that identification of an effect for evaluation requires consideration of existing data sets first and only if these are insufficient for biological risk assessment should additional testing be considered. This table is not a checklist for testing (see 6.5.1 and 6.6).

b For some medical devices, it can be appropriate to evaluate additional biological effects other than those indicated for evaluation in this table, for example medical devices manufactured from a novel material or intended for use in certain sensitive populations (e.g. young infants, pregnant women, immunocompromised individuals).

c It is also important to consider if there are specific circumstances which require evaluation of other biological effects (see 6.5.10) or consideration of other factors (see 6.5.11).

d For some medical devices with indirect contact, it is important to consider local effects after tissue contact if there is potential exposure to particulates (see 6.5.11.7).

e Evaluations for systemic toxicity should reflect the duration of use (see 6.5.5).

6.4.4.5 Medical devices in contact with either breached or compromised surfaces (skin or mucosal membranes) or internal tissues other than circulating blood

Medical devices in contact with either breached or compromised surfaces (skin or mucosal membranes) or internal tissues other than circulating blood are:

— surface contacting medical devices having direct contact or indirect contact with breached or compromised surfaces (skin or mucosal membranes);

EXAMPLE 1

Dressings or healing devices, occlusive patches for ulcers, burns and granulation tissue and wound cleansing and irrigation devices.

— medical devices with direct contact or indirect contact with bone, dentin, internal soft tissues or organs;

EXAMPLE 2

Medical devices that contact tissue, bone or pulp/dentin including laparoscopes, arthroscopes, draining systems, dental filling materials, skin staples, wires and biopsy needles.

EXAMPLE 3

Medical devices that have indirect contact by serving as conduits to store or deliver fluids to bone or tissues, including tubing sets used for irrigation, infusion or drainage.

EXAMPLE 4

Medical devices that contact bone including orthopaedic pins, plates, joint prostheses, bone cements and intraosseous devices.

EXAMPLE 5

Medical devices that contact soft tissues and tissue fluids including pacemakers, drug delivery devices, neuromuscular sensors and simulators, replacement tendons, breast implants, artificial larynxes, subperiosteal implants, ligation clips and intrauterine devices that do not achieve their primary function by chemical activity.

For these medical devices, the biological effects listed in Table 3 shall be evaluated.

Table 3
Biological effects for medical devices in contact with either breached or compromised surfaces (skin or mucosal membranes) or internal tissues other than circulating blood
Duration Biological effecta b c
Cytotoxicity (6.5.2) Sensitization (6.5.3) Irritation (6.5.4) Systemic toxicity (6.5.5)e Local effects after tissue contact (6.5.6)d Genotoxicity (6.5.7) Carcinogenicity (6.5.8) Haemocompatibility (6.5.9)
A – limited (≤24 h) E E E E        
B – prolonged (>24 h to 30 d) E E E E E E    
C – long-term (>30 d) E E E E E E E  

Key

E  biological effects to be evaluated

a It is important to emphasize that identification of an effect for evaluation requires consideration of existing data sets first and only if these are insufficient for biological risk assessment should additional testing be considered. This table is not a checklist for testing (see 6.5.1 and 6.6).

b For some medical devices it can be appropriate to evaluate additional biological effects other than those indicated for evaluation in this table, for example medical devices manufactured from a novel material or intended for use in certain sensitive populations (e.g. young infants, pregnant women, immunocompromised individuals).

c It is also important to consider if there are specific circumstances which require evaluation of other biological effects (see 6.5.10) or consideration of other factors (see 6.5.11).

d For some medical devices with indirect contact, it is important to consider local effects after tissue contact if there is potential exposure to particulates (see 6.5.11.7).

e Evaluations for systemic toxicity should reflect the duration of use (see 6.5.5).

6.4.4.6 Medical devices in contact with circulating blood

Medical devices in contact with circulating blood are medical devices having direct contact or indirect contact with circulating blood, including medical devices that store blood for human use or act as conduits to deliver fluids to or from blood.

EXAMPLE 1

Solution administration sets, extension sets, transfer sets and blood administration sets.

EXAMPLE 2

Temporary pacemaker electrodes, oxygenators, haemoadsorbents and immunoadsorbents.

EXAMPLE 3

Medical devices with components that are internal and external to the body, such as intravascular catheters or delivery systems, extracorporeal oxygenator tubing and accessories, dialysers, and dialysis tubing and accessories.

EXAMPLE 4

Components of pacemaker electrodes and leads within the cardiovascular system, artificial arteriovenous fistulae, heart valves, vascular grafts, prostheses and stents, internal drug-delivery catheters, vascular and cardiac implants, and ventricular assist devices.

This category is not intended to include devices only in contact with bleeding tissue (e.g. a hernia repair graft), which should be categorized as having contact with internal tissues (see 6.4.4.5).

For medical devices in contact with circulating blood, the biological effects listed in Table 4 shall be evaluated.

Table 4
Biological effects for medical devices in contact with circulating blood
Duration Biological effecta b c
Cytotoxicity (6.5.2) Sensitization (6.5.3) Irritation (6.5.4) Systemic toxicity (6.5.5)f Local effects after tissue contact (6.5.6) Genotoxicity (6.5.7) Carcinogenicity (6.5.8) Haemocompatibility (6.5.9)
A – limited (≤24 h) E E E E   Ee   E
B – prolonged (>24 h to 30 d) E E E E Ed E   E
C – long-term (>30 d) E E E E Ed E E E

Key

E biological effects to be evaluated

a It is important to emphasize that identification of an effect for evaluation requires consideration of existing data sets first and only if these are insufficient for biological risk assessment should additional testing be considered. This table is not a checklist for testing (see 6.5.1 and 6.6).

b For some medical devices, it can be appropriate to evaluate additional biological effects other than those indicated for evaluation in this table, for example medical devices manufactured from a novel material or intended for use in certain sensitive populations (e.g. young infants, pregnant women, immunocompromised individuals).

c It is also important to consider if there are specific circumstances which require evaluation of other biological effects (see 6.5.10) or consideration of other factors (see 6.5.11).

d Only considered for medical devices having direct contact with other tissues in addition to circulating blood contact as indirect contact is addressed in the haemocompatibility evaluation.

e Only for medical devices used in extracorporeal circuits or implantable medical devices.

f Evaluations for systemic toxicity should reflect the duration of use (see 6.5.5).

6.5 Biological effects for evaluation

6.5.1 Overall approach

This subclause provides guidance on the approach to evaluation of each separate biological effect identified and on circumstances which can require modifications in the approach to biological evaluation.

For a specified medical device, all biological effects identified in the relevant table shall be addressed in the biological evaluation and the approach taken shall be justified and documented.

Table 1, Table 2, Table 3 and Table 4 provide a framework for the biological evaluation of a medical device. Where these tables indicate a biological effect for evaluation, existing data sets relevant to that effect should be reviewed to determine if they are adequate to estimate biological risks, or whether additional information or testing is required to obtain reliable biological risk estimates (see 6.9).

If there are device-specific standards that include specific requirements or recommendations regarding biological evaluation or potential biological harms to be assessed, these should be considered in developing the approach to evaluation.

Rationales for changes to the requirements for evaluation of some biological effects included in Tables 1, Table 2, Table 3 and Table 4, compared to the previous version of this document, are provided in Annex B.

6.5.2 Cytotoxicity

Cytotoxicity evaluation is an initial screen for the presence of potential adverse biological effects from the medical device. Therefore, all medical devices having direct contact or indirect contact with the body, irrespective of contact duration, shall be evaluated for potential for cytotoxicity.

NOTE 1

A non-cytotoxic determination in this initial evaluation would not rule out the need for evaluation of other identified biological effects.

NOTE 2

Cytotoxicity data are assessed in relation to other biocompatibility data and the intended use of the product. Any cytotoxic effect can be of concern. However, it is primarily an indication of potential for in vivo toxicity and the device cannot necessarily be determined to be unsuitable for a given clinical application based solely on cytotoxicity data (see ISO 10993-5).

6.5.3 Sensitization

Exposure to constituents that are sensitizers can result in the induction of sensitization or the elicitation of an allergic response in sensitized individuals. Therefore, all medical devices having direct contact or indirect contact with the body, irrespective of contact duration, shall be evaluated for sensitization.

NOTE

Additional information on allergic responses is contained in 6.5.10.2 on immunotoxicity.

6.5.4 Irritation

Irritation can occur either because of the presence of an irritant constituent in or on the medical device, or as a result of a physical interaction, for example, from contact with particles released from a device or from physical interactions with the surface of a medical device. Therefore, all medical devices with direct contact or indirect contact with the body, irrespective of contact duration, shall be evaluated for irritation.

For example, components with indirect contact with blood (e.g. infusion systems) can introduce irritants into the blood stream. These components should be addressed as part of the biological risk assessment.

6.5.5 Systemic toxicity

Medical devices which can introduce constituents or particulates into systemic circulation, lymphatic system or cerebrospinal fluid shall be evaluated for the biological risk of harm from systemic toxicity.

The evaluation of systemic toxicity shall address the potential for adverse biological responses in tissues throughout the body. This evaluation shall include organ systems relevant to the clinical use of the medical device and other organs where chemical-specific toxicities can be relevant.

The duration of body contact is an important determinant of systemic toxicity and more extensive evaluation is required as duration of body contact increases (e.g. acute, sub-acute, sub-chronic and chronic). The approach to the evaluation for systemic toxicity shall be justified and documented.

EXAMPLE

Medical devices in long-term body contact are evaluated for acute, sub-acute, sub-chronic and chronic systemic toxicity.

Unless manufactured from novel materials or intended for use in certain sensitive populations, the evaluation for systemic toxicity is not usually required for medical devices:

— in contact with intact skin for any duration; or

— in limited duration contact with mucosal tissue.

However, if other factors apply to the medical device (see 6.5.11), an evaluation for systemic toxicity can be necessary.

Systemic toxicity can be evaluated in a toxicological risk assessment of chemical information. Any toxicological risk assessment shall be conducted in accordance with the requirements of ISO 10993-17 unless otherwise justified.

However, when testing is determined to be necessary, information obtained from comprehensive implantation assessments can also be appropriate to evaluate systemic toxicity if sufficient animals, dosing and timepoints are included and assessed. It is not always necessary to perform separate studies for acute, sub-acute, sub-chronic and chronic systemic toxicity. Such investigations, if appropriately designed and justified, can also combine evaluation of reproductive toxicity and local effects after tissue contact.

6.5.6 Local effects after tissue contact

If a medical device has direct contact with specific tissues or organs, then the local tissue responses to the medical device (including particulates) on those tissues or organs should be evaluated.

The duration of contact is important in the evaluation of local effects after tissue contact and more extensive evaluation is required as duration of body contact increases.

Unless manufactured from novel materials or intended for use in certain sensitive populations, an evaluation for local effects after tissue contact is not usually required for medical devices:

— in contact with intact skin for any duration; or

— in limited duration contact with any tissue including circulating blood.

An evaluation for local effects after tissue contact is required for medical devices in prolonged or long-term contact with tissues, except intact skin.

However, if other factors apply to the medical device (see 6.5.11), an evaluation for local effects can be necessary.

When testing is determined to be necessary, studies can be designed to estimate risks of biological harms related to local effects as well as to assess medical device performance or other biological effects such as systemic toxicity and haemocompatibility. Further information is contained in ISO 10993-6.

6.5.7 Genotoxicity

Genotoxicity evaluation is carried out to determine if constituents are present which can result in adverse effects related to genotoxicity. Duration of contact is important in determining the need for a genotoxicity assessment.

Unless manufactured from novel materials or intended for use in certain sensitive populations, an evaluation for genotoxicity is not usually required for medical devices:

— in contact with intact skin for any duration; or

— in limited duration contact with tissues, other than implantable medical devices with circulating blood contact, or medical devices that are part of extracorporeal circuits.

However, if other factors apply to the medical device (see 6.5.11), an evaluation for genotoxicity can be necessary.

Evaluation for genotoxicity is required for medical devices in:

— prolonged or long-term contact with all tissues (including circulating blood) except intact skin;

— limited duration contact with circulating blood, if implantable or part of an extracorporeal circuit.

However, if the medical device contains constituents known to be genotoxic, the potential for relevant biological harms should be considered in the biological risk assessment irrespective of the type or duration of body contact. Further information is contained in ISO 10993-3.

Genotoxicity can be evaluated in a toxicological risk assessment of chemical information. Any toxicological risk assessment shall be conducted in accordance with the requirements of ISO 10993-17 unless otherwise justified.

6.5.8 Carcinogenicity

Carcinogenicity can be of concern in medical devices resulting from:

— genotoxic mechanisms such as exposure to genotoxic constituents; or

— non-genotoxic mechanisms such as induction of cell proliferation or chronic inflammation[16][17][18].

Unless manufactured from novel materials or intended for use in certain sensitive populations, evaluation for carcinogenicity is not usually required for medical devices in contact with intact skin. However, if other factors (see 6.5.11) apply to the medical device, an evaluation for carcinogenicity can be necessary.

Medical devices in long term contact with any other tissues, including circulating blood shall be evaluated for carcinogenicity.

In vivo carcinogenicity tests use a large number of animals and are usually of limited sensitivity and relevance to exposure from medical devices. In the absence of any significant cancer risk, it is rare for carcinogenicity tests to be considered appropriate for medical devices. Chemical information and subsequent toxicological risk assessment can provide a basis for the evaluation of carcinogenicity. If the medical device contains constituents known to be carcinogenic, the potential for these harms should be considered in the biological risk assessment irrespective of the type or duration of body contact. Further information is contained in ISO 10993-3. Any toxicological risk assessment shall be conducted in accordance with the requirements of ISO 10993-17 unless otherwise justified.

6.5.9 Haemocompatibility

Devices having direct contact or indirect contact with circulating blood (irrespective of contact duration) can cause adverse interactions with the blood such as haemolysis and thrombosis and shall be evaluated for haemocompatiblity.

Haemocompatibility is affected by constituents and by the surface characteristics of the medical device, including the impact on blood flow. Therefore, understanding of relevant chemical and physical characteristics is important for blood contacting devices and shall be addressed in the biological risk assessment. Further information is contained in ISO 10993-4.

6.5.10 Other biological effects

6.5.10.1 Consideration of other effects

In addition to the specific biological effects described in Table 1, Table 2, Table 3 and Table 4, other biological effects relevant to the medical device shall be considered.

The following are specific examples.

6.5.10.2 Immunotoxicity

Immunotoxicity evaluation shall be considered based on:

— the physical or chemical properties of the materials of manufacture and data from sources that are suggestive of immunotoxicological effects, or

— if the immunogenic potential of any of the constituents is unknown.

Understanding of immunotoxicity is a developing science. ISO/TS 10993-20 provides information on the evaluation of immunotoxicology.

6.5.10.3 Neurotoxicity

Medical devices or their constituents that have direct contact or indirect contact with tissues of the central or peripheral nervous system or cerebrospinal fluid shall be evaluated for potential neurotoxicity. This evaluation should consider potential local neurotoxicity (local effects on the central or peripheral nervous tissue in close proximity to the medical device) as well as potential systemic neurotoxicity (systemic neurological effects not localized to the medical device)[19].

6.5.10.4 Reproductive and developmental toxicity

Reproductive and developmental toxicity shall be evaluated for medical devices:

— containing materials with a known reproductive or developmental toxicity or novel materials,

— indicated for relevant sensitive populations (e.g. foetuses, and pregnant or lactating women), or

— where there is the potential for local presence of medical device materials or constituents in the reproductive organs.

Evaluations shall consider the potential effects of medical devices, materials or their constituents on reproductive function, embryonic development (e.g. from exposure to teratogens), and prenatal and early postnatal development. These biological effects can be addressed with a biological risk assessment including chemical identification of impurities, extractable or leachable constituents, patient exposure to these constituents, weight of evidence and mode of action information, if available.

The evaluation for reproductive and developmental toxicity is not usually required for medical devices in contact with intact skin. However, if other factors (see 6.5.11) apply to the medical device, an evaluation for reproductive and developmental toxicity can be necessary.

Chemical information and subsequent toxicological risk assessment can provide a basis for the evaluation of the presence of potential reproductive or developmental toxicants. If the medical device contains constituents known to cause such adverse effects, the potential for these harms shall be considered in the biological risk assessment irrespective of the type or duration of body contact. Further information is contained in ISO 10993-3.

6.5.10.5 Material-mediated pyrogenicity

Pyrogenicity occurs when certain chemical or biological constituents of medical devices induce a febrile response. Chemically induced material-mediated pyrogenicity is rare and has been shown to occur only in the case of a small number of substances, which are not normally found as constituents.[20] Pyrogenicity can occur when medical devices are contaminated with biological or pharmaceutical pyrogens.

Medical devices containing constituents that have previously elicited a pyrogenic response or chemical entities where the pyrogenic potential is unknown (e.g. use of a novel material) should be evaluated for material-mediated pyrogenicity.

In many cases, it can be sufficient to present a rationale that;

— the composition of the finished medical device (including processing aids and manufacturing contaminants) does not present a significant biological risk of material-mediated pyrogenicity, and

— there are adequate controls to prevent pyrogenic contamination during manufacture.

Further information on material-mediated pyrogenicity is provided in ISO 10993-11: 2017.

NOTE

The evaluation for the presence of bacterial endotoxins and other microbial pyrogens is outside of the scope of this document and is addressed by ISO 11737-3.

6.5.11 Other factors to be considered

6.5.11.1 Consideration of other factors

The biological risk can be influenced by certain other factors. If any of these apply, the approach taken shall be justified and documented.

Other factors include (but are not limited to) those listed in 6.5.11.2 to 6.5.11.9.

6.5.11.2 Medical devices with very brief tissue contact

For medical devices that have very brief contact with the body, typically for less than one minute (e.g. lancets, hypodermic needles, capillary tubes), it is possible to prepare a written justification regarding the negligible potential for biological harm. For products with repeat use during a single 24 h period, the total exposure should be considered.

However, for medical devices that can leave materials in contact with tissues after the medical device is removed (e.g. coatings, lubricants) a more detailed biological evaluation is required.

6.5.11.3 Low risk intact skin contacting medical devices

A medical device or component which meets the following conditions is considered to have low biological risk if:

— contact is only with intact skin during normal use of the device, and

— the skin-contacting components are made from materials with a history of safe use in consumer products or medical devices with a similar type of contact and similar or longer duration of contact.

For these medical devices, a written justification shall be documented to include the materials, relevant prior use, any identified biological risks (e.g. from clinical use of similar medical devices), and why these biological risks would not apply for the medical device under consideration.

EXAMPLE

External surfaces and housings of equipment, patient examination tables, power cables and control interfaces such as keyboards, buttons, switches, dials and touch screens.

6.5.11.4 Absorption through the skin or mucosal membranes

Medical devices such as creams, drops or other topical preparations can result in substantial absorption of constituents through the skin or mucosal membranes. In this case, the biological evaluation shall consider the potential for systemic exposure to constituents released from the device and whether evaluation for systemic effects is required.

6.5.11.5 Materials that change during clinical use

If a medical device is intended to change during its clinical use, such as materials that polymerize or degrade in situ, the evaluation shall take into account all the different device states. For example, for an absorbable adhesive intended to polymerize in situ, the different device states would include starting components, intermediate reaction products, the fully polymerized material and subsequent degradation products.

6.5.11.6 Material degradation

The impact of degradation whether intended or not, shall be evaluated if:

— the medical device is designed to be absorbable or degradable,

— consideration of the finished medical device composition indicates that degradation products can be released during use, or

— physical degradation (including wear) can occur during use.

NOTE 1

Degradation can include chemical breakdown, corrosion or physical breakdown. Degradation products include both constituents and particulates (see 6.5.11.7).

NOTE 2

Further information on degradation related to mechanical wear can be found in device-specific standards (e.g. ISO 21535), as well as ISO/TR 10993-22.

In such cases, degradation products and the parameters that affect the rate and extent of degradation shall be documented. For medical devices intended to degrade, the expected mechanism of degradation shall also be documented.

EXAMPLE 1

Wear debris from total joint endoprostheses.

EXAMPLE 2

Release of constituents or particles from incomplete additive manufacturing or incomplete polymerisation of resins.

Specific guidance on degradation testing is provided in 6.8.3.

6.5.11.7 Particulates

If a medical device can release particulate materials into the tissues (e.g. from manufacturing residuals, through degradation or wear processes or by any other means) then the impact on biological safety of those particulate materials shall be considered in the biological evaluation.

For biological risk assessment of particulates, the following local and systemic effects shall be considered:

— effects related to physical characteristics; and

— effects related to the release of substances from the particulates.

NOTE

ISO/TS 10993-19 provides information on physico-chemical, morphological and topographic characterization of materials. ISO/TR 10993-22 provides specific information on nanomaterials.

6.5.11.8 Toxicokinetics

The purpose of conducting toxicokinetic studies is to evaluate the absorption, distribution, metabolism and excretion (ADME) of a constituent.

The evaluation of toxicokinetics shall be considered:

  1. a) if the medical device is designed to be absorbable,
  2. b) if the medical device is a long-term contact implant, and degradation or significant corrosion is known or likely,
  3. c) if the medical device is likely or known to release substantial quantities of potentially toxic or reactive degradation products or constituents that can be released into the body during clinical use,
  4. d) if substantial quantities of nano-objects are likely or known to be released from a medical device into the body during clinical use, or
  5. e) for drug and device combination products.

In this context, substantial quantities of constituents are amounts that can result in a risk of biological harm.

The release of degradation products or other constituents from metals, alloys and ceramics is usually too low to justify toxicokinetic studies, unless the material is designed to degrade.

Toxicokinetic studies are not required if the levels of clinical exposure at the expected rates of release of degradation products or other constituents have been demonstrated to be safe. Clinical experience or toxicological data can be used to support this approach.

Specific guidance on toxicokinetics studies is provided in 6.8.4.

6.5.11.9 Presence of hazardous constituents

If material information indicates that one or more constituents with known toxicities are present in a medical device at levels that can be toxic, the biological evaluation shall address the associated biological risk, regardless of the categorization (see 6.3).

6.6 Gap analysis

6.6.1 General

A gap analysis is a review of the available information to determine if the information is sufficient to carry out a biological risk assessment according to this document.

The gap analysis is first conducted following the initial data gathering. The gap analysis determines if the available information is sufficient either to demonstrate biological equivalence (see 6.7) or to proceed to biological risk estimation (see 6.9). If the information is not sufficient, additional information shall be gathered or generated, and the gap analysis repeated.

Additional information can include physical or chemical characterization, review of data from clinical use, or biological testing of the finished medical device.

The gap analysis shall be documented.

6.6.2 Medical devices evaluated using previous versions of this document

This document shall not be used to mandate re-testing of medical devices that are already on the market and have established and acceptable safety profiles. Instead, the historical information shall be reviewed and documented to determine whether any differences (e.g. change(s) in the medical device design or use since the prior review, additional biological effects, related literature and adverse events) will impact the conclusion of biological safety for the medical device under evaluation. A history of safe clinical use, that includes information relevant to the biological effects under consideration, can be sufficient such that no further biological evaluation is needed. However, if any of the changes described in production and post-production activities (see Clause 10) occur, an evaluation of the biological risks related to the change shall be performed using the current edition of this document.

If no additional biological effects are applicable (see 6.4.4), the update of prior biological evaluation documentation is not required. This shall be justified and documented as part of the review process.

6.7 Biological equivalence

Biological equivalence of a medical device can be demonstrated if the following are shown to be sufficiently similar to another medical device for which there are existing data demonstrating biological safety:

— type and duration of body contact;

— composition;

— physical characteristics or mechanical properties relevant to biological safety;

— manufacturing processes (including packaging and sterilization, if applicable);

— patient population; and

— size or quantity (number or amount) of material or medical device.

Sufficient similarity is where any differences do not result in any new or increased biological risks.

EXAMPLE

A constituent in the comparator application is present at a lower level in the new medical device or is present at a higher level which is still below an appropriate safety threshold.

If biological equivalence is claimed, the demonstration of equivalence shall be documented and justified based on following steps (see Figure 3).

— Chemical equivalence: situation where the chemical characteristics, including the composition and processing of the two materials or medical devices are similar such that differences in chemical characteristics do not result in additional or different biological risks (see ISO 10993-18 and ISO 10993-17).

— Physical equivalence: situation where the physical characteristics, including the configuration, morphology, topography and tribology, of two materials or medical devices are similar, such that the differences in physical characteristics do not result in additional or different biological risks (see ISO/TS 10993-19).

— Material equivalence: situation where two materials or medical devices demonstrate chemical and physical equivalence.

— Contact equivalence: situation where the clinical use, including patient population of two materials or medical devices is similar, such that the biological effects relevant to the evaluation are the same.

— Biological equivalence: situation where two materials or medical devices demonstrate material and contact equivalence.

Figure 4 describes the process for determining biological equivalence between two medical devices or materials.

If biological equivalence is demonstrated, then biological safety is established for the new or modified medical device.

Figure 3
Biological equivalence relationship map
Figure 4
Biological equivalence process map

6.8 Testing

6.8.1 General principles

If there are insufficient existing data to demonstrate safety, it can be necessary to carry out testing to obtain reliable estimates of biological risks and to complete the biological evaluation.

The following shall apply when testing of medical devices is considered necessary as part of the overall risk management process.

  1. a) Testing shall be performed on the finished medical device, or representative samples from the finished device or materials processed in the same manner as the finished medical device (including sterilization, if applicable). A representative sample can be a worst case taken from a medical device family. The choice of representative sample shall be justified and documented. For example, sample selection should account for process variation or changes that can occur during the medical device life cycle.
  2. b) For medical devices with multiple components that are tested separately due to different categorisations (see 6.4), interactions between the components shall be considered (e.g. simulated deployment prior to testing of an implanted component).
  3. c) If the medical device includes non-contacting components, the non-contacting components should not be included in the test article, unless justified.
  4. d) If the medical device includes both implanted and non-implanted components, and testing is carried out using an extract, the implanted components should not be combined in the same test article with the non-implanted components to prevent dilution of the implanted components, unless justified.
  5. e) The choice of test procedures shall consider:

    — type, duration, frequency and conditions of exposure to or contact with the medical device;

    — chemical and physical characteristics of the finished medical device;

    — toxicological activity of any known constituents in the formulation of the finished medical device;

    — the ratio of device surface area to recipient body size and mass;

    — animal welfare provisions (see 4.4); and

    — that in vivo biological tests are not justifiable where valid information related to released constituents and degradation products (including particulates) have been addressed by other means and the constituents have a known and acceptable toxicity profile (see 4.4).

  6. f) If extracts of the medical devices are prepared, the solvents and conditions of extraction used should be appropriate to the use of the finished medical device, including consideration for the potential loss of constituents during extraction. Extracts for biological testing shall be prepared in accordance with ISO 10993-12 and extracts for analytical chemistry testing shall be prepared in accordance with ISO 10993-18. Whenever possible, the extraction conditions should represent an exaggeration of use conditions.
  7. g) Positive and negative controls should be used where appropriate.

Testing shall be carried out in accordance with appropriate laboratory quality assurance controls [e.g. ISO/IEC 17025, OECD Principals of Good Laboratory Practice (GLP)]. Where test methods not detailed in the ISO 10993 series are used in the biological evaluation, these tests shall be sensitive, precise and accurate.

Where prior testing is considered during initial data gathering (see 6.6), evaluation of that testing shall be conducted and any differences relative to the applicable standards shall be justified and documented.

6.8.2 Biological, physical and chemical testing

If testing is needed, the selection of tests shall be based on device category as determined in accordance with 6.4. A rationale for the testing strategy, as well as for test selection, shall be documented in the biological evaluation plan (see Clause 5). Test protocols, reports and original data, complete to the extent that an independent analysis can be made, shall be evaluated by competent experts and shall be retained.

Any testing shall be performed in accordance with applicable requirements of other parts of the ISO 10993 series as listed in Table 5, unless otherwise justified.

If the characteristics of the device or its use requires a non-standardized test to be used, a scientific justification for the selected alternate test methods shall be documented.

Additional information on testing for materials mediated pyrogenicity is provided in ISO/TR 21582.

Table 5
ISO documents for biological, chemical and physical testing
Applicable document Test
ISO 10993-5 Cytotoxicity
ISO 10993-23 Irritation
ISO 10993-10 Sensitization
ISO 10993-11 Systemic toxicity
ISO 10993-6 Local effects after tissue contact
ISO 10993-3 Genotoxicity, carcinogenicity, reproductive and developmental toxicity
ISO 10993-4 Haemocompatibility
ISO 10993-6, ISO 10993-11 Neurotoxicity
ISO 10993-12 Preparation of samples for biological testing and use of primary reference materials
ISO 10993-18 Chemical characterization
ISO/TS 10993-19 Physical characterization
ISO 10993-7 Ethylene oxide residuals
ISO 10993-9, ISO 10993-13,
ISO 10993-14, ISO 10993-15
Identification and quantification of degradation products
ISO 10993-16 Toxicokinetic study design for degradation products and leachables

6.8.3 Degradation testing

If degradation testing is necessary (see 6.5.11.6), the parameters affecting the rate and extent of degradation shall be described and documented.

Expected degradation mechanisms should be simulated in vitro to estimate the rates of degradation and release of potentially toxic constituents to estimate the exposure. It is also possible that in vivo tests will be required to assess degradation of a material.

In vivo degradation tests can be unnecessary if an in vitro to in vivo comparison has been previously demonstrated and in vitro degradation studies show that only the probable products of degradation are present in the predicted quantities and produced at a rate similar to those that have a history of safe clinical use. When particulate degradation products are generated, tests can be unnecessary if the particulates are present in a quantity and physical state, i.e. size distribution and shape, similar to those with a history of safe clinical use or sufficient degradation data already exist relevant to the constituents and degradation products generated in the intended use.

A general framework for degradation tests is given in ISO 10993-9.

Specific in vitro degradation tests for polymers, ceramics and metals (and alloys) are described in ISO 10993-13, ISO 10993-14 and ISO 10993-15, respectively.

ISO/TS 37137-1 provides information on the evaluation of absorbable medical devices.

When particulate degradation products are present in the form of nanomaterials, tests can be designed considering ISO/TR 10993-22.

6.8.4 Toxicokinetic studies

If toxicokinetic studies are determined to be necessary (see 6.5.11.8), they shall be conducted in accordance with ISO 10993-16.

Where appropriate, theoretical degradation processes shall be investigated using in vitro experiments (e.g. with tissue, homogenates or cells) prior to in vivo toxicokinetic studies. This is for animal welfare reasons as given in ISO 10993-2, and to determine probable rather than possible degradation products.

Specific considerations for toxicokinetic studies with nanomaterials are given in ISO/TR 10993-22.

6.9 Biological risk estimation

Biological risk estimation is an input to determine biological safety. Biological risk estimation requires consideration of severity and probability of occurrence of harm.

For each biologically hazardous situation, the associated biological risk(s) shall be estimated using the data gathered or generated in accordance with applicable parts of Clause 6. Available information from published standards, literature, test data and clinical evidence can be used to estimate biological risk (see 6.1).

The biological risk can be estimated by assigning qualitative or quantitative values to the severity and probability of occurrence of biological harm. These values can be based on qualitative judgement or quantitative calculation, and they are combined to reach the biological risk estimate (for further information, see ISO/TR 24971:2020). The approach used for qualitative or quantitative estimation of the biological risk shall be documented.

The level of uncertainty inherent in the risk analysis shall be taken into account. If there is significant uncertainty due to limitations in relevant data, the risk estimate can be based on reasonable worst-case assumptions.

7 Biological risk evaluation

The identified and estimated biological risks shall be evaluated against the acceptability criteria defined in the biological evaluation plan (see Clause 5).

Examples of acceptability criteria comparisons can include:

— biological testing (e.g. cytotoxicity, irritation, sensitization) meets the acceptability criteria specified in the biological evaluation plan;

— degradation of an absorbable medical device does not elicit adverse local or systemic effects;

— toxicological risk assessment results in a margin of safety above a specified level derived according to ISO 10993-17; and

— the levels of exposure to residual ethylene oxide are below the thresholds specified in ISO 10993-7.

Biological safety is established when acceptability criteria are met.

If one or more biological risks do not meet the acceptability criteria, then the following shall be considered:

  1. a) refinement of biological risk estimates which can include:

    — the improvement in constituent quantification in accordance with ISO 10993-18;

    — the selection of more appropriate tolerable intake values based on additional gathered literature in accordance with ISO 10993-17; or

    — biological testing (see 6.8).

  2. b) implementation of biological risk controls (see Clause 8).

The risks shall then be re-evaluated for acceptability with the following action outcomes:

— If re-evaluation determines that the risks are now acceptable, biological safety is established and no further action is needed.

— If estimates of one or more biological risks are found to still be unacceptable or a new unacceptable biological risk is identified following risk control measures, then these biological risks shall be further controlled to the greatest extent feasible in accordance with Clause 8.

— When one or more biological risks still do not meet the acceptability criteria, a benefit-risk assessment shall be conducted of the residual biological risk in accordance with ISO 14971.

8 Biological risk control

If estimates of one or more biological risks do not meet the acceptability criteria, then these biological risks shall be controlled (reduced) to the greatest extent feasible.

Where biological risk control is necessary, one or more of the following risk control options shall be used, in the priority order listed (see ISO 14971: 2019):

  1. a) inherently safe design and manufacture (e.g. substitute with a less toxic material);
  2. b) protective measures in the medical device itself (e.g. application of a barrier around a toxic material) or in the manufacturing process (e.g. addition of a cleaning step to remove a toxic constituent);
  3. c) information for safety and, where appropriate, training to users (e.g. contraindications for susceptible individuals, warnings about biological risks from reasonably foreseeable misuse or a risk of biological harm).

If risk control measures are implemented, they shall be evaluated for acceptability (see Clause 7).

If the overall residual biological risk, after risk control measures, is judged acceptable, the manufacturer shall inform users of significant residual biological risks and shall include the necessary information in the accompanying documentation in order to disclose those residual biological risks (see ISO 14971: 2019).

9 Biological evaluation report

After all information has been obtained in accordance with the biological evaluation plan and any biological risk control measures have been implemented and verified, the findings and conclusions shall be documented in the biological evaluation report.

The report shall be prepared by competent personnel (e.g. lead author, contributors). The evidence of competence shall be documented.

The biological evaluation report shall document, as applicable:

  1. a) the criteria for determining the acceptability (see Clause 5) of the medical device or medical device family for the intended use or for reasonably foreseeable misuse, in line with the biological evaluation plan;
  2. b) conformity with the biological evaluation plan (see Clause 5);
  3. c) how each biological effect applicable to the medical device has been addressed;
  4. d) the interpretation and assessment of data including historical information, relevance to the finished medical device and the results of relevant physical, chemical or biological testing;
  5. e) the rationale for:

    — the selection of tests or waiving of tests, or both, or

    — the use of non-standard test methods not described in any part of the ISO 10993 series;

  6. f) the outcomes of the biological risk evaluation (see Clause 7) and associated biological risk estimations (see 6.9);
  7. g) any biological risk control (see Clause 8) measures applied;
  8. h) identified residual biological risks and overall biological safety conclusions for the medical device.

10 Production and post-production activities

Biological evaluation is conducted within a total life cycle risk management framework. A system shall be established and maintained to actively collect, document and review information relevant to the biological evaluation of the medical device in the production and post-production phases. Therefore, the biological evaluation shall be reviewed whenever changes to design, production processes or emerging post-market information can impact biological safety.

The biological evaluation documentation shall be reviewed and updated as required if any of the following occur:

— user or patient feedback or other post-market information relevant to biological safety which indicates that previously unidentified biological hazards exist, or biological risk estimates need to be adjusted;

— a change in specification, formulation, processing or source of any material with direct contact or indirect contact;

— a change in processing, primary packaging or sterilization of the medical device;

— a change in the manufacturer's instructions or expectations concerning storage (e.g. changes in either shelf life or transport, or both); or

— a change in intended use of the medical device (e.g. change resulting in a different body contact, duration of use or different patient populations).

The review shall consider:

— the changes which have occurred or new information obtained;

— a change in standards referenced in the biological evaluation documentation (see 6.6.2);

— the impact on biological hazards, potential biological harms and biological risk estimates; and

— the omission of or need for additional evaluations, including any additional testing.

The review shall be documented and include:

— an overall statement of conclusion as to biological safety, and

— a scientific justification for the decisions to conduct, not conduct or limit the scope of additional evaluations or testing.

Annex A Material selection and characterization to support the biological evaluation of a medical device

A.1 Background information

The selection and characterization of materials used in a medical device are important aspects of the biological evaluation of a medical device.

A.2 Material selection

Material selection and biological risk analysis are integral components of the design process for medical devices and are considered in parallel throughout product development.

The selection of materials plays a crucial role in controlling the biological risk and, when approached in a systematic way, allows the collection of relevant data. In line with ISO 14971, criteria to define the acceptable biological risk are established at the start of the design process.

When selecting materials for a medical device, their suitability is established based on the intended purpose, taking into account their chemical, toxicological, physical, electrical, morphological and mechanical properties. In addition to material selection, manufacturing processes (including sterilization), packaging, transportation or storage can impact finished medical device biological safety. These additional factors are considered in the biological risk assessment.

The biological evaluation is planned and performed to demonstrate the achievement of specified criteria for safety based on outputs of the biological risk analysis or history of use of the same material.

A.3 Characterization

A.3.1 Chemical characterization

From a practical perspective, chemical characterization information (see ISO 10993-18) are most useful in a biological evaluation when:

— a limited number of constituents are changed in a medical device;

— sufficient information cannot be obtained from the vendor (e.g. proprietary issues cannot be resolved, see A.3.3);

— analytical chemistry studies are feasible to conduct; or

— toxicity data are available for the constituents.

A.3.2 Use of chemical characterization data in a biological evaluation

Information on constituents of the medical device is useful in identifying potential biological risks.

EXAMPLE 1

If a constituent is known to produce nephrotoxic effects, attention to this biological hazard is needed when conducting acute or sub-chronic toxicity assessments.

EXAMPLE 2

If a constituent is known to be a carcinogen, attention to this biological hazard is needed when conducting a toxicological risk assessment.

Evaluation of such information can be used to focus the biological evaluation strategy to address the relevant biological effects for the finished medical device.

Chemical characterization data can also be useful for biological risk estimation. If data are available on the rate at which a constituent is released from the medical device under conditions that simulate the in-use environment, and if sufficient data are available to derive a relevant safety threshold or chemical specific limit, then it is possible to compare the dose received to the relevant threshold or limit in assessing probability of adverse effects.

Analytical chemistry testing is not always necessary if there is sufficient information for biological risk estimation. Additionally, in some cases analytical chemistry testing is not appropriate (e.g. when the medical device is incompatible with extraction solvents or other methods are available to evaluate the biological effect). For example, carcinogenicity evaluation can be performed using available genotoxicity and carcinogenicity information, including from the literature, relevant to the materials of construction and manufacturing processes.

A.3.3 Proprietary material information

Where the necessary data (e.g. complete formulation and manufacturing information) are not available to a manufacturer because of confidentiality of proprietary information, enquiries can be made with the materials supplier as to the availability of biological evaluations for the material, which can be relevant to the proposed application. In some cases, it is possible to manage confidentiality of proprietary formulations by means of separate submission by the owner of the biological evaluation data to an independent assessor or regulatory agency. These data can then be referenced in a regulatory submission by the medical device manufacturer and confidentially reviewed by the relevant conformity assessment body or regulatory agency in conjunction with the medical device submission review. If these methods to obtain relevant proprietary information are not successful, testing can be needed as described in the rest of this document.

A.3.4 Physical characterization

Physical characteristics (e.g. configuration, particle size, porosity, surface texture) can have a substantial effect on biological interactions with the medical device and can impact safety. It is important to consider such aspects as part of the biological evaluation. If there are insufficient data available from literature or other sources to estimate biological risks, then it is possible that further investigations using appropriate functional models or other investigations of the effect of physical characteristics will be necessary. Examples include:

— evaluation of configuration on blood flow and haemocompatiblity;

— evaluation of porosity on tissue ingrowth;

— evaluation of release of wear particles on local and distant tissue responses; or

— evaluation of surface texture (topography) on cell adhesion, phenotypic expression and growth.

For particulates and nanomaterials (when used in medical devices), physical characterization is necessary, as described for nanomaterials in ISO/TR 10993-22.

A.3.5 Effects of manufacturing processes

It is important to consider the effect of manufacturing conditions on materials and the use of additives or presence of contaminants. In general, to support biological safety, testing is performed on test samples processed and sterilised (if applicable) in equivalent ways to the finished medical device. Where the material processing differs from that used to produce test articles to generate the test data, a justification is required to explain why the differences are not significant to the determination of biological safety. Particular aspects to consider include:

— processes which can cause either bulk or surface changes in materials properties e.g. moulding, surface treatment, welding or machining;

— intended additives or processing aids, such as catalysts, antioxidants, pigments, surface treatments and others;

— potential process contaminants (e.g. cleaning agents, disinfection agents, sterilization agents, etching agents, mould release agents, cutting fluids and particles, and machine contaminants such as lubricants or residues from the manufacture of device components made from different materials); and

— degradation during manufacturing and processing.

Annex B Rationale for key changes in the biological effects listed in Tables 1 to 4

B.1 Background information

This Annex summarizes the key changes between the fifth edition of this document (i.e. ISO 10993-1:2018) and this edition of this document in device categorization and in biological effects to be considered for evaluation and provides the rationale for these changes.

B.2 Changes to device contact categories

In this document, the number of contact categories has been simplified to four groups of devices in contact with:

— intact skin (see 6.4.4.3 and Table 1);

— intact mucosal membranes (see 6.4.4.4 and Table 2);

— breached or compromised surfaces or internal tissues other than blood (see 6.4.4.5 and Table 3);

— blood (see 6.4.4.6 and Table 4).

Externally communicating medical devices are no longer a separate contact category and are not separately referenced in this document as the term externally communicating was challenging to translate in some regions. The recommended biological effects for externally communicating medical devices for the relevant type and duration of contact are now included either:

— in the category of medical devices in contact with breached or compromised surfaces or internal tissues other than circulating blood, including medical devices contacting internal tissue, bone or dentin, and breached or compromised skin (see 6.4.4.5 and Table 3); or

— in the blood contacting category, which includes, medical devices having direct contact or indirect contact with blood including medical devices that store blood or act as conduits to deliver fluids to or from blood, such as medical devices with components that are internal and external to the body (see 6.4.4.6 and Table 4).

B.3 Changes to the presentation of systemic toxicity

The prior edition of this document (i.e. ISO 10993-1:2018) included acute, subacute, sub-chronic and chronic toxicity in Table A.1. This current edition of this document includes a general category of systemic toxicity and indicates in the table footnotes that the evaluations for systemic toxicity should reflect the duration of use.

B.4 Changes in requirements for genotoxicity and carcinogenicity evaluation

B.4.1 Overview

The prior edition of this document (i.e. ISO 10993-1:2018) did not include consistent recommendations for genotoxicity and carcinogenicity evaluation for medical device categories with the same types and durations of contact. This current edition of this document has introduced some changes to address these inconsistencies as indicated in B.4.2 to B.4.3.

B.4.2 Genotoxicity evaluation of prolonged contact (>24 h to 30 d) medical devices

A genotoxicity evaluation has been added for medical devices contacting mucosal, breached or compromised surfaces and having indirect contact with blood (see Table 2, Table 3 and Table 4). Genotoxicity assessment is now applicable for all medical devices with prolonged contact, except for medical devices with only intact skin contact.

The development of cancer is considered a multistage process.[21][22] Each additional exposure to a genotoxic constituent increases the risk for cancer development. When the exposure is systemic, a genotoxic constituent can reach multiple tissues increasing the risk. Therefore, a genotoxicity evaluation is applicable for all prolonged contact medical devices requiring a systemic toxicity evaluation.

A genotoxicity evaluation can be used as a potential screen for carcinogens.[23] Therefore, if the outcome of the genotoxicity evaluation indicates that a genotoxic constituent is present, an evaluation for genotoxic carcinogenicity is applicable.

The development of cancer by a non-genotoxic mechanism requires long-term exposure and is not expected to occur in prolonged contact.[16] Therefore, for prolonged contact, an evaluation for non-genotoxic carcinogenicity is considered unnecessary.

B.4.3 Carcinogenicity evaluation of long-term contact (>30 d) medical devices

A carcinogenicity evaluation is applicable for all long-term contact medical devices where chronic toxicity is evaluated. Therefore, a carcinogenicity evaluation has been added for medical devices with long-term mucosal contact (see Table 2).

B.5 Omission of material-mediated pyrogenicity from Table 1 to Table 4

Material-mediated pyrogenicity has been removed from Table 1, Table 2, Table 3 and Table 4 because this biological effect is rare and has been shown to occur only in the case of a small number of constituents. Guidance on material-mediated pyrogenicity is now located in 6.5.10.5 as part of other biological effects for evaluation.

Bibliography

  • [1] ISO 18562 (all parts) Biocompatibility evaluation of breathing gas pathways in healthcare applications
  • [2] ISO 7405 Dentistry — Evaluation of biocompatibility of medical devices used in dentistry
  • [3] ISO 11737-3 Sterilization of health care products — Microbiological methods — Part 3: Bacterial endotoxin testing
  • [4] ISO 22442 (all parts) Medical devices utilizing animal tissues and their derivatives
  • [5] ISO 13485:2016 Medical devices — Quality management systems — Requirements for regulatory purposes
  • [6] ISO/TR 10993-22 Biological evaluation of medical devices — Part 22: Guidance on nanomaterials
  • [7] ISO 14155 Clinical investigation of medical devices for human subjects — Good clinical practice
  • [8] ISO/TR 24971:2020 Medical devices — Guidance on the application of ISO 14971
  • [9] ISO/TS 10993-20 Biological evaluation of medical devices — Part 20: Principles and methods for immunotoxicology testing of medical devices
  • [10] ISO/TR 21582 Pyrogenicity — Principles and methods for pyrogen testing of medical devices
  • [11] ISO/TS 37137-1 Biological evaluation of absorbable medical devices — Part 1: General requirements
  • [12] ISO 21535 Non-active surgical implants — Joint replacement implants — Specific requirements for hip-joint replacement implants
  • [13] ISO/IEC 17025 General requirements for the competence of testing and calibration laboratories
  • [14] ISO/IEC Guide 63:2019 Guide to the development and inclusion of aspects of safety in International Standards for medical devices
  • [15] Organisation for Economic Co-operation and Development (OECD). (1998). OECD Principles on Good Laboratory Practice. OECD Series on Principles of Good Laboratory Practice and Compliance Monitoring, Number 1. Paris: OECD Publishing.
  • [16] Wolf D. C., Cohen S. M., Boobis A. R., Dellarco V. L., Fenner-Crisp P. A., Moretto A., Pastoor T. P., Schoeny R. S., Seed J. G., Doe J. E. (2019) Chemical Carcinogenicity Revisited 1: A unified theory of carcinogenicity based on contemporary knowledge, Regulatory Toxicology and Pharmacology, 103: 86-92
  • [17] Hernández L., van Steeg H., Luijten M., van Benthem J. Mechanisms of non-genotoxic carcinogens and importance of a weight of evidence approach. Mutation Research/Reviews in Mutation Research 2009,682 (2–3), 94-109. Available from: https://doi.org/10.1016/j.mrrev.2009.07.002
  • [18] Smith M., Guyton K., Gibbons C., Fritz J., Portier C., Rusyn I., DeMarini D., Caldwell J., Kavlock R., Lambert P., Hecht S., Bucher J., Stewart B., Baan R., Cogliano V., Straif K. Key Characteristics of Carcinogens as a Basis for Organizing Data on Mechanisms of Carcinogenesis. Environ Health Persp 2016, 124 (6), 713-721. Available from: https://doi.org/10.1289/ehp.1509912
  • [19] Guidance Document Redbook 2000: IV.C.10. Neurotoxicity Studies, 2020. Available from: https://www.fda.gov/regulatory-information/search-fda-guidance-documents/redbook-2000-ivc10-neurotoxicity-studies
  • [20] Borton L. and Coleman K. (2025) Material-mediated pyrogens in medical devices: Myth or reality? ALTEX. Available from: https://doi.org/10.14573/altex.2504231
  • [21] Owins D., Wei C., Smart R. (1999) A Multihit, Multistage Model of Carcinogenesis, Carcinogenesis, 20 ; 1837-1844.
  • [22] Abel E. L., Angel J. M., Kiguchi K., DiGiovanni J. (2009) Multi-stage chemical carcinogenesis in mouse skin: Fundamentals and applications 4(9): 1350–1362.
  • [23] IARC Scientific Publications, No. 165. Baan R. A., Stewart, B. W. Straif K., editors. Lyon (FR): International Agency for Research on Cancer, 2019. Available from: https://publications.iarc.fr/Book-And-Report-Series/Iarc-Scientific-Publications
  • [24] ISO 10993 (all parts) Biological evaluation of medical devices