ISO / ISO 10993-17:2023/Amd 1:2025 Biological evaluation of medical devices — Part 17: Toxicological risk assessment of medical device constituents — AMENDMENT 1

ISO 10993-17:2023/Amd 1:2025 Biological evaluation of medical devices — Part 17: Toxicological risk assessment of medical device constituents — AMENDMENT 1

Foreword

ISO (the International Organization for Standardization) is a worldwide federation of national standards bodies (ISO member bodies). The work of preparing International Standards is normally carried out through ISO technical committees. Each member body interested in a subject for which a technical committee has been established has the right to be represented on that committee. International organizations, governmental and non-governmental, in liaison with ISO, also take part in the work. ISO collaborates closely with the International Electrotechnical Commission (IEC) on all matters of electrotechnical standardization.

The procedures used to develop this document and those intended for its further maintenance are described in the ISO/IEC Directives, Part 1. In particular, the different approval criteria needed for the different types of ISO documents should be noted. This document was drafted in accordance with the editorial rules of the ISO/IEC Directives, Part 2 (see www.iso.org/directives).

ISO draws attention to the possibility that the implementation of this document may involve the use of (a) patent(s). ISO takes no position concerning the evidence, validity or applicability of any claimed patent rights in respect thereof. As of the date of publication of this document, ISO had not received notice of (a) patent(s) which may be required to implement this document. However, implementers are cautioned that this may not represent the latest information, which may be obtained from the patent database available at www.iso.org/patents. ISO shall not be held responsible for identifying any or all such patent rights.

Any trade name used in this document is information given for the convenience of users and does not constitute an endorsement.

For an explanation of the voluntary nature of standards, the meaning of ISO specific terms and expressions related to conformity assessment, as well as information about ISO's adherence to the World Trade Organization (WTO) principles in the Technical Barriers to Trade (TBT), see www.iso.org/iso/foreword.html.

This document was prepared by Technical Committee ISO/TC 194, Biological and clinical evaluation of medical devices, in collaboration with the European Committee for Standardization (CEN) Technical Committee CEN/TC 206, Biological and clinical evaluation of medical devices, in accordance with the Agreement on technical cooperation between ISO and CEN (Vienna Agreement).

A list of all parts in the ISO 10993 series can be found on the ISO website.

Any feedback or questions on this document should be directed to the user’s national standards body. A complete listing of these bodies can be found at www.iso.org/members.html.

Biological evaluation of medical devices — Part 17: Toxicological risk assessment of medical device constituents — AMENDMENT 1

 

Clause 2, Normative references

Add a footnote to the following reference:

ISO 10993-18:20201, Biological evaluation of medical devices — Part 18: Chemical characterization of medical device materials within a risk management process

 

1 As amended by ISO 10993-18:2020/Amd 1:2022.

 

5.2, Figure 2

In the first diamond, remove ‘or’ in ‘or ‘TQ>TSL’.

 

In the second box, remove ‘of’ in ‘Apply of TTC (7.2)’.

 

7.1, NOTE 1

Move NOTE 2 above NOTE 1 and renumber the notes accordingly.

 

Clause B.1, sixth paragraph

In the sixth paragraph, replace the second bullet with the following:

  — TQ of each constituent present in or on or released from the medical device when the TQmax represents cumulative exposure dose over the duration that the medical device contacts the body.

 

Clause B.1, seventh paragraph

Replace the seventh paragraph with the following:

Toxicological screening limits shall not apply to medical devices used for a prolonged or long-term duration in preterm or very young infants, or the following types of constituents:

 

Clause B.2, first paragraph

In the first paragraph, replace the first sentence with the following:

Toxicological screening limits are based on mutagenic TTC values applicable to the duration of exposure to any single constituent.

 

In the first paragraph, replace the second sentence with the following:

The toxicological screening limit shall be calculated using Formula (B.1).

 

Clause B.2, Formula (B.1)

Under Formula (B.1), replace the description of ‘D’ with the following:

D is the minimum assumed duration of exposure expressed in days (d) for the time period of assumed exposure considered for the constituent in accordance with ISO 10993-1:2018.

 

Clause B.2, Table B.1

In the table header for the TTC column, replace the unit ‘g/d’ with ‘µg/d’.

 

Clause B.3, Formula (B.2)

Replace the description of ‘SF’ for Formula (B.2) with the following:

SF is a unitless value that represents the ratio of the maximum number or quantity (e.g. cm2, g or ml) of medical devices that are in contact with the body during the period of assumed exposure to the number or quantity of either medical device(s) used in the extraction study or from information gathered from formulation and manufacturing process data (e.g. cm2, g or ml).

 

Clause B.3, EXAMPLES 1 to 5

In EXAMPLES 1 to 5, replace each description of ‘1’ in the TSL formulae with the following:

1 is the minimum number of days that applies to ≤30 d assumed exposure time period (see Table B.1).

 

Clause B.3, EXAMPLE 6

In EXAMPLE 6, replace the description of ‘10’ with the following:

10 is the SF where 500 cm2 (50 cm2 × 10 devices) is the maximum device surface area, in cm2, that is in contact with the body divided by the 50 cm2 device surface area extracted.

 

Clause E.1, b) and c)

In the first sentence for each item, replace ‘6.2.1’ with ‘6.2.2’ and replace ‘total quantity’ with ‘maximum total quantity (TQmax, in µg, the cumulative exposure dose)’.

 

Clause E.2, Tables E.2 and E.3

In the table headers for the HQr.k. column of Table E.2 and Table E.3, replace the unit "μg" with "μg/d".

 

E.3.1, fifth paragraph

Replace the fifth paragraph with the following:

When the TQ of a constituent is obtained by extraction and quantification methods that are conservative relative to the medical device intended use, the TQ of each reportable constituent may be assumed to represent a cumulative exposure dose (TQmax) and is adequate for calculating a worst-case estimated exposure dose, EEDmax.

 

Clause F.3

In the second sub-list item of the thirtheenth list item, replace ‘TQ’ with ‘TQmax’.

 

Clause F.4, Table F.3

In the table header of the third column, replace ‘TQ extracted’ with ‘TQmax’.

In the table header of the fifth, and sixth columns, replace ‘TQ’ with ‘TQmax’.

 

In footnote c, replace ‘total quantity’ with ‘maximum total quantity’.